QRS Detection on Heart Rate Variability Readings using Two Moving Average Methods

Heart Rate Variability or the average deviation between heartbeats in humans is influenced by the autonomic nervous system control of heart function. Monitoring HRV is necessary to diagnose the underlying pathophysiology of hypertension, optimize treatment modalities for hypertensive patients with signs of autonomic dysfunction, and predict cardiovascular events in the heart. This study focused on providing an overview of QRS complex detection for heart rate variability or HRV reading using the Two Moving Average method in detecting heart in humans. In addition, current research also determine QRS complex detection for heart rate variability reading by adding a window size feature, then create a QRS Complex detection tool for HRV reading using the Two Moving Average method by adding a window size feature. Furthermore, another aim of this study is to know the FFT signal results in order to see the frequency of each ECG signal generated by the patient. In this study, the use of the Two Moving Average method or moving average makes it easier to find the R peak-to-peak signal, so the heartbeats reading is easier as well. In this study, QRS complex signal detection was performed using lead II pickups using the Two Moving Average method, which was used as a filter or attenuator of unsought signals such as P and T signals in ECG signals. In this case, this method is recommended for detecting patients with high P and T signal values. This was achieved by evaluating and studying each change in window size, an algorithm that uses an equation with two different window widths to generate signal features and detection thresholds, allowing it to adapt to various changes in QRS and noise levels.

Authors

Institutions

Publication Details

Journal
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Published
2025-08-17
DOI
https://doi.org/10.35882/ijeeemi.v5i1.165
Primary Topic
Heart Rate Variability and Autonomic Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

QRS Detection on Heart Rate Variability Readings using Two Moving Average Methods

I Dewa Gede Hari Wisana, Andjar Pudji, Ayu Nissa Berlianri Rizhky, Sima Des
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Heart Rate Variability and Autonomic Control
article

QRS Detection on Heart Rate Variability Readings using Two Moving Average Methods

I Dewa Gede Hari Wisana, Andjar Pudji, Ayu Nissa Berlianri Rizhky, Sima Des
article en

Abstract

Heart Rate Variability or the average deviation between heartbeats in humans is influenced by the autonomic nervous system control of heart function. Monitoring HRV is necessary to diagnose the underlying pathophysiology of hypertension, optimize treatment modalities for hypertensive patients with signs of autonomic dysfunction, and predict cardiovascular events in the heart. This study focused on providing an overview of QRS complex detection for heart rate variability or HRV reading using the Two Moving Average method in detecting heart in humans. In addition, current research also determine QRS complex detection for heart rate variability reading by adding a window size feature, then create a QRS Complex detection tool for HRV reading using the Two Moving Average method by adding a window size feature. Furthermore, another aim of this study is to know the FFT signal results in order to see the frequency of each ECG signal generated by the patient. In this study, the use of the Two Moving Average method or moving average makes it easier to find the R peak-to-peak signal, so the heartbeats reading is easier as well. In this study, QRS complex signal detection was performed using lead II pickups using the Two Moving Average method, which was used as a filter or attenuator of unsought signals such as P and T signals in ECG signals. In this case, this method is recommended for detecting patients with high P and T signal values. This was achieved by evaluating and studying each change in window size, an algorithm that uses an equation with two different window widths to generate signal features and detection thresholds, allowing it to adapt to various changes in QRS and noise levels.

Indonesian Journal of Electronics Electromedical Engineering and Medical InformaticsVol. 5(1)
Politeknik Kesehatan Kemenkes Semarang (ID), Eskom (South Africa) (ZA)
Openalex Percentile: Top 66%
Heart Rate Variability and Autonomic Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

QRS Detection on Heart Rate Variability Readings using Two Moving Average Methods — I Dewa Gede Hari Wisana, Andjar Pudji, et al. · Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics (2025) | TGRS Research Map | TGRS