Comparative Analysis of Water and Oil Media on Temperature Stability in PID Control-Based Digital Thermometer Calibrator

Digital thermometers are measuring instruments needed to perform temperature measurement actions and must be calibrated periodically according to standard measurement methods. The purpose of developing this tool is to add PID control to the calibration media where PID control aims to regulate the stability of the temperature setting to be achieved. This is achieved by studying and evaluating the effect of temperature stability on the heater and LM35DZ temperature sensor. This research method uses the Arduino Nanosystem for data processing and PID system control. The LM35DZ temperature sensor on the heater is regulated by a 2 Channel SSR module using a PID system then the temperature generated by the heater will be read by the LM35DZ and displayed on the LCD. The results of this study, digital thermometer calibrator measurements have been successfully carried out by comparing 3 digital thermometers with different brands, namely Omron 343F, Omron 245, and ThermoOne. The difference in error values in oil media is 3-4% and in water media 2-4% with the value of time stability in water media for 3-3.3 minutes and in oil media for 1-2 hours. With this comparison of calibration media, it is hoped that it can help in measuring temperature with better and more effective results. find methods, results, conclusions.

Authors

Institutions

Publication Details

Journal
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Published
2025-08-17
DOI
https://doi.org/10.35882/ijeeemi.v5i2.171
Primary Topic
Sensor Technology and Measurement Systems
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative Analysis of Water and Oil Media on Temperature Stability in PID Control-Based Digital Thermometer Calibrator

Bedjo Utomo, Anggara Trisna Nugraha, Andjar Pudji, Syaifudin Syaifudin et al.
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Sensor Technology and Measurement Systems
article

Comparative Analysis of Water and Oil Media on Temperature Stability in PID Control-Based Digital Thermometer Calibrator

Bedjo Utomo, Anggara Trisna Nugraha, Andjar Pudji, Syaifudin Syaifudin, Mochammad Sofyan
article en

Abstract

Digital thermometers are measuring instruments needed to perform temperature measurement actions and must be calibrated periodically according to standard measurement methods. The purpose of developing this tool is to add PID control to the calibration media where PID control aims to regulate the stability of the temperature setting to be achieved. This is achieved by studying and evaluating the effect of temperature stability on the heater and LM35DZ temperature sensor. This research method uses the Arduino Nanosystem for data processing and PID system control. The LM35DZ temperature sensor on the heater is regulated by a 2 Channel SSR module using a PID system then the temperature generated by the heater will be read by the LM35DZ and displayed on the LCD. The results of this study, digital thermometer calibrator measurements have been successfully carried out by comparing 3 digital thermometers with different brands, namely Omron 343F, Omron 245, and ThermoOne. The difference in error values in oil media is 3-4% and in water media 2-4% with the value of time stability in water media for 3-3.3 minutes and in oil media for 1-2 hours. With this comparison of calibration media, it is hoped that it can help in measuring temperature with better and more effective results. find methods, results, conclusions.

Indonesian Journal of Electronics Electromedical Engineering and Medical InformaticsVol. 5(2)
Universitas Negeri Surabaya (ID), Politeknik Kesehatan Kemenkes Semarang (ID), Poltekkes Kemenkes Surabaya
Openalex Percentile: Top 77%
Sensor Technology and Measurement Systems
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Comparative Analysis of Water and Oil Media on Temperature Stability in PID Control-Based Digital Thermometer Calibrator — Bedjo Utomo, Anggara Trisna Nugraha, et al. · Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics (2025) | TGRS Research Map | TGRS