Development of a Low-Cost and Efficient ECG devices with IIR Digital Filter Design

Measurement of biosignals such as electrocardiograph has the interpretation of noise from other signals. The noise can interfere with the measurement of the heart signal and make the measurement inaccurate, so the purpose of this study is to make a 6-Lead Electrocardiogram module with an Arduino-Based Digital Filter. By using a digital filter. The contribution of this research is the use of digital filters to eliminate noise in electrocardiograph signals. This research uses Infinite Impulse Filter digital filters such as Butterworth, Chebyshev I, Chebyshev II, and Elliptic in order 2, 4, 6, 8, and 10. The study was conducted by providing input from the Function Generator on Arduino which has been applied digital filters with Frequency with 0.5Hz – 100Hz cut-off. The instrument is compared with a factory electrocardiograph. Filter measurements using 460 input data. Butterworth filter with the greatest emphasis on order 8 frequency 0.5Hz produces an emphasis of -5.74298158 dB and a frequency of 100Hz produces an emphasis of -5.93529424 dB. The Chebyshev I filter has the greatest emphasis on order 6 frequency 0.5Hz producing an emphasis of -3.27104076 dB and on order 8 frequency 100Hz producing an emphasis of -5.08730424 dB. Chebyshev II filter the biggest emphasis on the order of frequency 0.5Hz produces a suppression of -44,66011104 dB and 80Hz frequency produces a suppression of -37,3653957 dB. Elliptic filters the greatest emphasis on order 6 frequency 0.5Hz produces an emphasis on -1.55429354 dB and 100Hz frequency on order 8 produces an emphasis on -2.2849115 dB. The results of this study can be implemented on a 6-Lead ECG module to eliminate noise or interference when tapping ECG signals.

Authors

Institutions

Publication Details

Journal
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
Published
2025-08-17
DOI
https://doi.org/10.35882/ijeeemi.v3i1.194
Citations
1
Primary Topic
ECG Monitoring and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a Low-Cost and Efficient ECG devices with IIR Digital Filter Design

I Dewa Gede Hari Wisana
1 citations
Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics
ECG Monitoring and Analysis
article

Development of a Low-Cost and Efficient ECG devices with IIR Digital Filter Design

I Dewa Gede Hari Wisana
article en
1 citations

Abstract

Measurement of biosignals such as electrocardiograph has the interpretation of noise from other signals. The noise can interfere with the measurement of the heart signal and make the measurement inaccurate, so the purpose of this study is to make a 6-Lead Electrocardiogram module with an Arduino-Based Digital Filter. By using a digital filter. The contribution of this research is the use of digital filters to eliminate noise in electrocardiograph signals. This research uses Infinite Impulse Filter digital filters such as Butterworth, Chebyshev I, Chebyshev II, and Elliptic in order 2, 4, 6, 8, and 10. The study was conducted by providing input from the Function Generator on Arduino which has been applied digital filters with Frequency with 0.5Hz – 100Hz cut-off. The instrument is compared with a factory electrocardiograph. Filter measurements using 460 input data. Butterworth filter with the greatest emphasis on order 8 frequency 0.5Hz produces an emphasis of -5.74298158 dB and a frequency of 100Hz produces an emphasis of -5.93529424 dB. The Chebyshev I filter has the greatest emphasis on order 6 frequency 0.5Hz producing an emphasis of -3.27104076 dB and on order 8 frequency 100Hz producing an emphasis of -5.08730424 dB. Chebyshev II filter the biggest emphasis on the order of frequency 0.5Hz produces a suppression of -44,66011104 dB and 80Hz frequency produces a suppression of -37,3653957 dB. Elliptic filters the greatest emphasis on order 6 frequency 0.5Hz produces an emphasis on -1.55429354 dB and 100Hz frequency on order 8 produces an emphasis on -2.2849115 dB. The results of this study can be implemented on a 6-Lead ECG module to eliminate noise or interference when tapping ECG signals.

Indonesian Journal of Electronics Electromedical Engineering and Medical InformaticsVol. 3(1)
Politeknik Kesehatan Kemenkes Semarang (ID), Eskom (South Africa) (ZA)
Openalex Percentile: Top 66%
ECG Monitoring and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Development of a Low-Cost and Efficient ECG devices with IIR Digital Filter Design — I Dewa Gede Hari Wisana · Indonesian Journal of Electronics Electromedical Engineering and Medical Informatics (2025) | TGRS Research Map | TGRS