Test–Retest Reliability and Inter-Scanner Reproducibility of Improved Spinal Diffusion Tensor Imaging

<b>Background/Objectives</b>: Spinal diffusion tensor imaging (sDTI) remains a challenging method for the selective evaluation of key anatomical structures, like pyramidal tracts (PTs) and dorsal columns (DCs), and for reliably quantifying diffusion metrics such as fractional anisotropy (FA), radial diffusivity (RD), mean diffusivity (MD), and axial diffusivity (AD). This prospective, single-center study aimed to assess the reproducibility, robustness, and reliability of an optimized axial sDTI protocol, specifically intended for long fiber tracts. <b>Methods</b>: We developed an optimized Stejskal-Tanner sequence for high-resolution, axial sDTI of the cervical spinal cord at 3.0 T. Using advanced standardized evaluation and post-processing methods, we estimated DTI values for PTs, DCs, and AHs at the level of the second cervical vertebra. Reliability was evaluated through repeated measurements in 16 healthy volunteers and by comparing results from two 3.0 T scanners (Magnetom Skyra and Magnetom Prisma, Siemens Healthineers, Erlangen, Germany). Reproducibility was assessed using paired t-tests, intraclass correlation coefficients (ICCs), Bland-Altman analysis, and coefficients of variation (CVs). <b>Results</b>: The optimized sDTI protocol demonstrated high consistency for FA between test-retest sessions and across scanners. For the Skyra, the DC region showed the highest reliability (average ICC = 0.858) followed by the PT region (average ICC = 0.789). On the Prisma, the PT region reached an average ICC of 0.854, with the DC region at 0.758. Pooled inter-scanner data indicated good-to-excellent agreement, particularly in the PT region (average ICC = 0.860). FA CVs remained low (<10%) across all regions and scanners. RD showed good-to-excellent ICC values for PTs and DCs (average ICC for Skyra 0.642 and 0.769 and 0.926 and 0.830 for Prisma, respectively) but showed a higher CV between 14.6 and 19.4% for these two scanners. <b>Conclusions</b>: Improved sDTI offers highly reproducible FA measurements for all metrics with scanner independence, supporting its potential as a robust tool for detecting and monitoring spinal cord pathologies.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-08-16
DOI
https://doi.org/10.3390/diagnostics15162057
Primary Topic
Advanced Neuroimaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Test–Retest Reliability and Inter-Scanner Reproducibility of Improved Spinal Diffusion Tensor Imaging

Georg Gohla, Tobias Lindig, Christer Ruff, Tim W. Rattay et al.
Diagnostics
Advanced Neuroimaging Techniques and Applications
article

Test–Retest Reliability and Inter-Scanner Reproducibility of Improved Spinal Diffusion Tensor Imaging

Georg Gohla, Tobias Lindig, Christer Ruff, Tim W. Rattay, Benjamin Bender, Stephan König, Uwe Klose, Ulrike Ernemann
article en

Abstract

<b>Background/Objectives</b>: Spinal diffusion tensor imaging (sDTI) remains a challenging method for the selective evaluation of key anatomical structures, like pyramidal tracts (PTs) and dorsal columns (DCs), and for reliably quantifying diffusion metrics such as fractional anisotropy (FA), radial diffusivity (RD), mean diffusivity (MD), and axial diffusivity (AD). This prospective, single-center study aimed to assess the reproducibility, robustness, and reliability of an optimized axial sDTI protocol, specifically intended for long fiber tracts. <b>Methods</b>: We developed an optimized Stejskal-Tanner sequence for high-resolution, axial sDTI of the cervical spinal cord at 3.0 T. Using advanced standardized evaluation and post-processing methods, we estimated DTI values for PTs, DCs, and AHs at the level of the second cervical vertebra. Reliability was evaluated through repeated measurements in 16 healthy volunteers and by comparing results from two 3.0 T scanners (Magnetom Skyra and Magnetom Prisma, Siemens Healthineers, Erlangen, Germany). Reproducibility was assessed using paired t-tests, intraclass correlation coefficients (ICCs), Bland-Altman analysis, and coefficients of variation (CVs). <b>Results</b>: The optimized sDTI protocol demonstrated high consistency for FA between test-retest sessions and across scanners. For the Skyra, the DC region showed the highest reliability (average ICC = 0.858) followed by the PT region (average ICC = 0.789). On the Prisma, the PT region reached an average ICC of 0.854, with the DC region at 0.758. Pooled inter-scanner data indicated good-to-excellent agreement, particularly in the PT region (average ICC = 0.860). FA CVs remained low (<10%) across all regions and scanners. RD showed good-to-excellent ICC values for PTs and DCs (average ICC for Skyra 0.642 and 0.769 and 0.926 and 0.830 for Prisma, respectively) but showed a higher CV between 14.6 and 19.4% for these two scanners. <b>Conclusions</b>: Improved sDTI offers highly reproducible FA measurements for all metrics with scanner independence, supporting its potential as a robust tool for detecting and monitoring spinal cord pathologies.

DiagnosticsVol. 15(16)
Christian-Albrechts-Universität zu Kiel (DE), University Hospital Schleswig-Holstein (DE), University of Tübingen (DE), University of Lübeck (DE)
Openalex Percentile: Top 70%
Advanced Neuroimaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.