Ultrasonic Selective Opening of Microcavities for Drug Delivery Microimplants

We present a fully passive, ultraminiaturized drug delivery microchip that enables wireless, on-demand release via ultrasonic actuation - a mechanism not previously demonstrated in implantable drug delivery systems. By eliminating the need for integrated power and control components, the device achieves sub-millimeter dimensions (<0.2 mm <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\times 0.5$</tex-math> </inline-formula> mm <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\times 1$</tex-math> </inline-formula> mm), enabling minimally invasive implantation in sensitive or hard-to-reach tissues. The system relies on ultrathin metallic membranes (<100<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\mu $</tex-math> </inline-formula>m wide, <200 nm thick), engineered to rupture selectively at predefined ultrasonic resonance frequencies. Membrane behavior is optimized through analytical and numerical modeling, and the devices are fabricated using standard, scalable MEMS microfabrication techniques. Selective actuation enables precise release of up to <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$2~\mu $</tex-math> </inline-formula>g of powder payload from individual microcavities in response to specific frequencies (360/420/580kHz), with membrane rupture occurring in <italic xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">in vivo</i> conditions at ultrasonic intensities within clinically safe limits (<150 mW/cm<italic xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <sup>2</sup> </i>). This platform represents a significant step toward precise, programmable drug delivery in anatomically constrained or delicate regions.[2025-0104]

Authors

Institutions

Publication Details

Journal
Journal of Microelectromechanical Systems
Published
2025-08-18
DOI
https://doi.org/10.1109/jmems.2025.3597789
Primary Topic
Microfluidic and Bio-sensing Technologies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Ultrasonic Selective Opening of Microcavities for Drug Delivery Microimplants

Niclas Roxhed, Theocharis Nikiforos Iordanidis, Göran Stemme, Argyris Spyrou
Journal of Microelectromechanical Systems
Microfluidic and Bio-sensing Technologies
article

Ultrasonic Selective Opening of Microcavities for Drug Delivery Microimplants

Niclas Roxhed, Theocharis Nikiforos Iordanidis, Göran Stemme, Argyris Spyrou
article en

Abstract

We present a fully passive, ultraminiaturized drug delivery microchip that enables wireless, on-demand release via ultrasonic actuation - a mechanism not previously demonstrated in implantable drug delivery systems. By eliminating the need for integrated power and control components, the device achieves sub-millimeter dimensions (<0.2 mm <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\times 0.5$</tex-math> </inline-formula> mm <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\times 1$</tex-math> </inline-formula> mm), enabling minimally invasive implantation in sensitive or hard-to-reach tissues. The system relies on ultrathin metallic membranes (<100<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\mu $</tex-math> </inline-formula>m wide, <200 nm thick), engineered to rupture selectively at predefined ultrasonic resonance frequencies. Membrane behavior is optimized through analytical and numerical modeling, and the devices are fabricated using standard, scalable MEMS microfabrication techniques. Selective actuation enables precise release of up to <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$2~\mu $</tex-math> </inline-formula>g of powder payload from individual microcavities in response to specific frequencies (360/420/580kHz), with membrane rupture occurring in <italic xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">in vivo</i> conditions at ultrasonic intensities within clinically safe limits (<150 mW/cm<italic xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <sup>2</sup> </i>). This platform represents a significant step toward precise, programmable drug delivery in anatomically constrained or delicate regions.[2025-0104]

Journal of Microelectromechanical SystemsVol. 34(5)
KTH Royal Institute of Technology (SE)
The Research Council
Good health and well-being
Openalex Percentile: Top 80%
Microfluidic and Bio-sensing Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.