Validation of blood pressure measurement using a smartwatch in patients with acute ischemic stroke
We aimed to validate the accuracy of blood pressure (BP) measurement using a smartwatch in patients with acute ischemic stroke. We compared 140 pairs of BP (n = 35) measurements acquired by a smartwatch (ASUS VivoWatch SP) with those measured by a sphygmomanometer (reference device). Differences between the smartwatch BP and reference BP measurements were compared. The validation procedure and criterion followed the consensus of the American National Standards Institute, Inc/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2018 and extended the standard to the specificities of cuffless devices in acute ischemic stroke population. The mean and standard deviation of the differences measured by a smartwatch and reference device were 1.8 ± 5.7 mm Hg in systolic BP and 0.7 ± 3.6 mm Hg in diastolic BP according to criterion 1. The mean and standard deviation of the differences measured by a smartwatch and reference device were 1.8 ± 5.6 mm Hg in systolic BP and 0.7 ± 3.6 mm Hg in diastolic BP according to criterion 2. The results both met the standards of the 2 criterion. The validation result did not differ between the paralytic and non-paralytic arms. The smartwatch with photoplethysmography sensors can provide accurate and reliable measurement of BP in acute ischemic stroke patients.
Authors
- Yu‐Hsuan Liu (ORCID: https://orcid.org/0000-0002-0113-3695)
- Yan Guo (ORCID: https://orcid.org/0000-0003-1282-5344)
- Li-Ching Chen (ORCID: https://orcid.org/0000-0003-4940-2791)
- Huan-Jan Lin (ORCID: https://orcid.org/0000-0003-2524-4928)
Institutions
- Kaohsiung Medical University (TW)
- Kaohsiung Medical University Chung-Ho Memorial Hospital (TW)
- E-Da Hospital (TW)
- Ministry of Health and Welfare (TW)
- I-Shou University (TW)
Publication Details
- Journal
- Medicine
- Published
- 2025-08-15
- DOI
- https://doi.org/10.1097/md.0000000000042899
- Primary Topic
- Hemodynamic Monitoring and Therapy
- Type
- article
- Field-Weighted Citation Impact
- 0.00