Introduction: Respiratory oscillometry (OSC) is a pulmonary function test that measures respiratory system impedance and is well-suited for young children. Previous studies have suggested that there may be differences between impedance values obtained on different instruments, however, no studies comparing instruments have been done in healthy children. Methods: We performed both impulse oscillometry (IOS) and airwave oscillometry (AOS), in random order, in a cohort of healthy children with no history of cardiopulmonary disease or recent illness. We used linear regression analysis and Bland Altman plots to compare results. Results: Eighty children ages 4 – 18 years completed study visits. Compared with AOS, the IOS mean and median values for resistance at 5 Hz (R5) were higher (p < 0.001), reactance at 5 Hz (X5) mean and median values were less negative (p < 0.001), and both the area under the reactance curve (AX) and resonant frequency (Fres) were lower (p < 0.001). Bland Altman analysis revealed proportional bias between the two systems. Conclusions: Our results in healthy children show differences between results obtained on different OSC instruments. This may be due to differences in the type of pressure signal produced by the instrument, or, in the case of IOS, may be related to a stress-relaxation response. Our results suggest that normal values obtained on one instrument may not be comparable to results obtained on another.

Authors

Institutions

Publication Details

Published
2025-08-18
DOI
https://doi.org/10.22541/au.175554196.62476180/v1
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Discrepancy between airwave and impulse oscillometry measurements of impedance in healthy children

Catherine Qiu, Clement L. Ren, Maureen B. Josephson, Julian Lewis Allen et al.
Non-Invasive Vital Sign Monitoring
article

Discrepancy between airwave and impulse oscillometry measurements of impedance in healthy children

Catherine Qiu, Clement L. Ren, Maureen B. Josephson, Julian Lewis Allen, Samuel Goldfarb, Sara B. DeMauro, Kristina Ziolkowski, Heather Boas, Tryce Scully, Laurie Travaglini, Katharine Tsukahara, Joseph M. McDonough
article en

Abstract

Introduction: Respiratory oscillometry (OSC) is a pulmonary function test that measures respiratory system impedance and is well-suited for young children. Previous studies have suggested that there may be differences between impedance values obtained on different instruments, however, no studies comparing instruments have been done in healthy children. Methods: We performed both impulse oscillometry (IOS) and airwave oscillometry (AOS), in random order, in a cohort of healthy children with no history of cardiopulmonary disease or recent illness. We used linear regression analysis and Bland Altman plots to compare results. Results: Eighty children ages 4 – 18 years completed study visits. Compared with AOS, the IOS mean and median values for resistance at 5 Hz (R5) were higher (p < 0.001), reactance at 5 Hz (X5) mean and median values were less negative (p < 0.001), and both the area under the reactance curve (AX) and resonant frequency (Fres) were lower (p < 0.001). Bland Altman analysis revealed proportional bias between the two systems. Conclusions: Our results in healthy children show differences between results obtained on different OSC instruments. This may be due to differences in the type of pressure signal produced by the instrument, or, in the case of IOS, may be related to a stress-relaxation response. Our results suggest that normal values obtained on one instrument may not be comparable to results obtained on another.

University of Minnesota (US), Children's Hospital of Philadelphia (US), University of Utah (US), Twin Cities Orthopedics (US), University of Pennsylvania (US)
Children's Hospital of Philadelphia, National Heart, Lung, and Blood Institute
Openalex Percentile: Top 80%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.