Evaluation of commercial point-of-care glucometers for detection and monitoring of neonatal hypoglycemia in resource-constrained settings

Accurate identification and treatment of hypoglycemia or low blood glucose is important for neonates, especially premature, small for gestational age, and intra-uterine growth restricted (“small and sick”) newborns. Existing point-of-care (POC) adult glucometers, intended for home management of adults or children with diabetes, are widely available and less costly than machines designed and validated for neonates, but they must be validated at the lower glucose levels relevant for neonates and at appropriate hematocrit levels before being used for neonates. We simulated neonatal blood by altering hematocrit (to 18%, 40%, and 55%) and glucose concentration (to 40 mg/dL and 150 mg/dL) of normal adult blood. We then used 11 POC glucometers to measure glucose levels in these samples and compared results to values measured with a clinical chemistry analyzer (YSI 2300). We report mean percent bias, mean absolute relative difference, and standard deviation for each tested condition and glucometer. We ranked the tested glucometers based on a combination of accuracy and precision for measuring neonatal glucose levels and consumable cost. From best-performing to worst, they are Group 1: StatStrip and StatStrip Xpress 2; Group 2: AccuChek Guide, AccuChek GuideMe, AccuChek Instant, and AccuChek Performa; Group 3: AccuChek Active, GlucoNavii; Group 4: HemoCue RT, Nipro Premier, and OneTouchUltra 2. This paper describes a simple, laboratory-based method to test glucometers across a wide range of hematocrit values and glucose concentrations and contributes up-to-date testing data on currently available POC glucometers at ranges relevant to neonatal use.

Authors

Institutions

Publication Details

Journal
BMC Pediatrics
Published
2025-08-16
DOI
https://doi.org/10.1186/s12887-025-05934-9
Primary Topic
Hyperglycemia and glycemic control in critically ill and hospitalized patients
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of commercial point-of-care glucometers for detection and monitoring of neonatal hypoglycemia in resource-constrained settings

Meaghan Bond, Chinyere Ezeaka, Elizabeth Molyneux, Rebecca Richards‐Kortum et al.
BMC Pediatrics
Hyperglycemia and glycemic control in critically ill and hospitalized patients
article

Evaluation of commercial point-of-care glucometers for detection and monitoring of neonatal hypoglycemia in resource-constrained settings

Meaghan Bond, Chinyere Ezeaka, Elizabeth Molyneux, Rebecca Richards‐Kortum, Elizabeth Asma, Z. Maria Oden, Joseph Peterson, Danica Kumara, Millicent Alooh, Lauren Oparah
article en

Abstract

Accurate identification and treatment of hypoglycemia or low blood glucose is important for neonates, especially premature, small for gestational age, and intra-uterine growth restricted (“small and sick”) newborns. Existing point-of-care (POC) adult glucometers, intended for home management of adults or children with diabetes, are widely available and less costly than machines designed and validated for neonates, but they must be validated at the lower glucose levels relevant for neonates and at appropriate hematocrit levels before being used for neonates. We simulated neonatal blood by altering hematocrit (to 18%, 40%, and 55%) and glucose concentration (to 40 mg/dL and 150 mg/dL) of normal adult blood. We then used 11 POC glucometers to measure glucose levels in these samples and compared results to values measured with a clinical chemistry analyzer (YSI 2300). We report mean percent bias, mean absolute relative difference, and standard deviation for each tested condition and glucometer. We ranked the tested glucometers based on a combination of accuracy and precision for measuring neonatal glucose levels and consumable cost. From best-performing to worst, they are Group 1: StatStrip and StatStrip Xpress 2; Group 2: AccuChek Guide, AccuChek GuideMe, AccuChek Instant, and AccuChek Performa; Group 3: AccuChek Active, GlucoNavii; Group 4: HemoCue RT, Nipro Premier, and OneTouchUltra 2. This paper describes a simple, laboratory-based method to test glucometers across a wide range of hematocrit values and glucose concentrations and contributes up-to-date testing data on currently available POC glucometers at ranges relevant to neonatal use.

BMC PediatricsVol. 25(1)
University of Lagos (NG), Kamuzu Central Hospital (MW), Kamuzu University of Health Sciences (MW), Rice University (US)
Bill and Melinda Gates Foundation, John D. and Catherine T. MacArthur Foundation, Rice University, Children's Investment Fund Foundation, Lemelson Foundation, Ting Tsung and Wei Fong Chao Family Foundation
Openalex Percentile: Top 71%
Hyperglycemia and glycemic control in critically ill and hospitalized patients
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.