Personalized dosimetry for Re-188-SIRT: a comparison of gafchromic film and commercial tools—are we delivering what we intent?

<i>Objective.</i>Personalized dosimetry in radiopharmaceutical therapy offers superior accuracy over generic methods, though absorbed dose (AD) estimates vary due to differing techniques. While Monte Carlo simulation (MCS) is precise but complex, gafchromic film dosimetry offers a simpler alternative with strong agreement to MCS for<i>β</i>-AD. We present a cost-effective, high-resolution gafchromic film dosimetry using a custom polymethyl methacrylate (PMMA) phantom for Rhenium-188. A Dosimetric Volume Unit (DVU) was defined and a conversion factor (J GBq<sup>-1</sup>) was used to estimate the AD to liver tumors of varying volumes via mono-compartmental method in<sup>188</sup>Re-Selective Intra-arterial Radionuclide Therapy (SIRT) patients. Results were compared with three commercial softwares: OLINDA/EXM® V2.0, MIRDcalc, and QThera.AI.<i>Approach.</i>A tissue-equivalent PMMA phantom incorporating a central cavity for EBT3 gafchromic film placement was developed for high-resolution dosimetry of<sup>188</sup>Re. Optical density measurements from gafchromic film exposed to varying<sup>188</sup>Re activities were utilized to quantify AD as a function of radioactivity. Percentage depth dose curves were generated using 25<i>µ</i>Ci of<sup>188</sup>Re for AD estimates at every 0.275 mm depth increments along the entire path length. Dose fall-off profiles were used to define the DVU, from which a conversion factor (J GBq<sup>-1</sup>) was derived and applied for patient-specific AD estimation using mono-compartmental method in 25<sup>188</sup>Re-SIRT patients. The results were compared with dosimetric outputs from three commonly-used softwares.<i>Main results.</i>A perfect linear correlation (<i>R</i><sup>2</sup>= 1) was observed between AD and radioactivity. In a homogeneous distribution, 1<i>µ</i>Ci of<sup>188</sup>Re delivered 3.6 Gy mm<sup>-3</sup>in the first half-life and a cumulative 7.3 Gy mm<sup>-3</sup>over ten half-lives, with a maximum tissue penetration of 3.7 mm. The DVU dimensions for<sup>188</sup>Re were estimated as 4.7 × 4.7 × 7.4 mm<sup>3</sup>, and the conversion factor was calculated as 34 J GBq<sup>-1</sup>. In 25 patients, the median AD calculated using mono-compartmental model was 77.5 Gy (Interquartile range IQR: 57.1-120.2 Gy), closely aligning with QThera.AI estimates [87.1 (IQR: 43.0-127.9) Gy;<i>p</i>= 0.968] and showing a strong correlation (<i>r</i>= 0.88). In contrast, MIRDcalc produced significantly lower ADs [47.6 (IQR: 29.4-84.0) Gy;<i>p</i>= 0.000] with a moderate correlation (<i>r =</i> 0.62), while OLINDA/EXM® V2.0 yielded the lowest values [13.5 (IQR: 7.0-24.) Gy;<i>p</i>= 0.000] and exhibited a weak correlation<i>(r</i>= 0.58).<i>Significance.</i>DVU based mono-compartmental method showed strong concordance with QThera.AI estimates, supporting its use for accurate, empirical AD calculation in<sup>188</sup>Re-SIRT, similar to<sup>90</sup>Y dosimetry. Given the high cost of commercial softwares, this low-cost approach may offer a practical alternative, promoting broader adoption of precision dosimetry, particularly in resource-limited settings.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-08-15
DOI
https://doi.org/10.1088/1361-6560/adfc22
Citations
1
Primary Topic
Radiopharmaceutical Chemistry and Applications
Type
article
Field-Weighted Citation Impact
1.36

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Personalized dosimetry for Re-188-SIRT: a comparison of gafchromic film and commercial tools—are we delivering what we intent?

Gaganpreet Singh, Jaya Shukla, Arun S. Oinam, Sejal Chopra et al.
1 citations
Physics in Medicine and Biology
Radiopharmaceutical Chemistry and Applications
1.36
article

Personalized dosimetry for Re-188-SIRT: a comparison of gafchromic film and commercial tools—are we delivering what we intent?

Gaganpreet Singh, Jaya Shukla, Arun S. Oinam, Sejal Chopra, Komalpreet Kaur, BhagwantRai Mittal
article en
1 citations

Abstract

<i>Objective.</i>Personalized dosimetry in radiopharmaceutical therapy offers superior accuracy over generic methods, though absorbed dose (AD) estimates vary due to differing techniques. While Monte Carlo simulation (MCS) is precise but complex, gafchromic film dosimetry offers a simpler alternative with strong agreement to MCS for<i>β</i>-AD. We present a cost-effective, high-resolution gafchromic film dosimetry using a custom polymethyl methacrylate (PMMA) phantom for Rhenium-188. A Dosimetric Volume Unit (DVU) was defined and a conversion factor (J GBq<sup>-1</sup>) was used to estimate the AD to liver tumors of varying volumes via mono-compartmental method in<sup>188</sup>Re-Selective Intra-arterial Radionuclide Therapy (SIRT) patients. Results were compared with three commercial softwares: OLINDA/EXM® V2.0, MIRDcalc, and QThera.AI.<i>Approach.</i>A tissue-equivalent PMMA phantom incorporating a central cavity for EBT3 gafchromic film placement was developed for high-resolution dosimetry of<sup>188</sup>Re. Optical density measurements from gafchromic film exposed to varying<sup>188</sup>Re activities were utilized to quantify AD as a function of radioactivity. Percentage depth dose curves were generated using 25<i>µ</i>Ci of<sup>188</sup>Re for AD estimates at every 0.275 mm depth increments along the entire path length. Dose fall-off profiles were used to define the DVU, from which a conversion factor (J GBq<sup>-1</sup>) was derived and applied for patient-specific AD estimation using mono-compartmental method in 25<sup>188</sup>Re-SIRT patients. The results were compared with dosimetric outputs from three commonly-used softwares.<i>Main results.</i>A perfect linear correlation (<i>R</i><sup>2</sup>= 1) was observed between AD and radioactivity. In a homogeneous distribution, 1<i>µ</i>Ci of<sup>188</sup>Re delivered 3.6 Gy mm<sup>-3</sup>in the first half-life and a cumulative 7.3 Gy mm<sup>-3</sup>over ten half-lives, with a maximum tissue penetration of 3.7 mm. The DVU dimensions for<sup>188</sup>Re were estimated as 4.7 × 4.7 × 7.4 mm<sup>3</sup>, and the conversion factor was calculated as 34 J GBq<sup>-1</sup>. In 25 patients, the median AD calculated using mono-compartmental model was 77.5 Gy (Interquartile range IQR: 57.1-120.2 Gy), closely aligning with QThera.AI estimates [87.1 (IQR: 43.0-127.9) Gy;<i>p</i>= 0.968] and showing a strong correlation (<i>r</i>= 0.88). In contrast, MIRDcalc produced significantly lower ADs [47.6 (IQR: 29.4-84.0) Gy;<i>p</i>= 0.000] with a moderate correlation (<i>r =</i> 0.62), while OLINDA/EXM® V2.0 yielded the lowest values [13.5 (IQR: 7.0-24.) Gy;<i>p</i>= 0.000] and exhibited a weak correlation<i>(r</i>= 0.58).<i>Significance.</i>DVU based mono-compartmental method showed strong concordance with QThera.AI estimates, supporting its use for accurate, empirical AD calculation in<sup>188</sup>Re-SIRT, similar to<sup>90</sup>Y dosimetry. Given the high cost of commercial softwares, this low-cost approach may offer a practical alternative, promoting broader adoption of precision dosimetry, particularly in resource-limited settings.

Physics in Medicine and BiologyVol. 70(17)
Post Graduate Institute of Medical Education and Research (IN), Panjab University (IN)
Indian Council of Medical Research
Good health and well-being
Openalex Percentile: Top 17%
Radiopharmaceutical Chemistry and Applications
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.