<i>Objective.</i>Spot-scanning proton arc therapy (SPArc) enables dynamic dose delivery through continuous gantry rotation, offering potential improvements in dose conformity for complex targets. However, dynamic delivery introduces susceptibility to gantry position errors (GPEs), which may degrade the treatment accuracy, particularly in highly precise and conformal techniques. This study aims to quantitatively assess the dosimetric impact of systematic and random GPEs during the dynamic SPArc delivery.<i>Method.</i>Twolve clinical cases, including brain stereotactic radiosurgery (SRS), lung, liver, and prostate cancer, were evaluated under simulated systematic and random GPEs with ±0.5°, ±.1°, ±1.5°, ±2° per 180° arc scenarios. SPArc plans were generated using a published algorithm implemented in RayStation through scripting and simulated the treatment via a validated IBA DynamicARC® system model. Virtual machine logfiles reconstructed control points and spot positions, incorporating GPEs through modified gantry velocities. Target coverage (D98) and gamma passing ratios (GPR) under 3%/3 mm and 2%/2 mm criteria were analyzed.<i>Results.</i>Under the 3%/3 mm criterion, GPR exceeded 98% for all cases; liver, lung, and prostate plans remained >99% across GPE magnitudes, with lung, liver, and prostate plans exceeding 99.4% even at a 2° systematic error. Brain SRS cases were most sensitive; under a 2%/2 mm criterion, the GPR dropped to 89.52 ± 4.74% for a 2° systematic error. In contrast, lung, liver, and prostate plans maintained GPRs above 93% under the same conditions. The impact of random GPEs was minimal, with all plans achieving GPRs greater than 96% even for the strictest 2%/2 mm criterion.<i>Significance.</i>This study provides the first comprehensive quantitative evaluation of dose perturbation caused by GPE in the dynamic SPArc treatment delivery. Systematic GPEs introduce more significant dose deviations compared to random errors, with the impact varying by cases, disease sites and error magnitude. These findings could guide machine quality assurance protocols and site-specific error tolerance thresholds for future clinical implementation.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-08-15
DOI
https://doi.org/10.1088/1361-6560/adfc21
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Investigate gantry position error effect on dynamic spot-scanning proton arc (SPArc) therapy

Gang Liu, Xuanfeng Ding, Lewei Zhao, Xiangkun Xu et al.
Physics in Medicine and Biology
Radiation Therapy and Dosimetry
article

Investigate gantry position error effect on dynamic spot-scanning proton arc (SPArc) therapy

Gang Liu, Xuanfeng Ding, Lewei Zhao, Xiangkun Xu, Xiaoqiang Li, Xiaoda Cong, P. Liu
article en

Abstract

<i>Objective.</i>Spot-scanning proton arc therapy (SPArc) enables dynamic dose delivery through continuous gantry rotation, offering potential improvements in dose conformity for complex targets. However, dynamic delivery introduces susceptibility to gantry position errors (GPEs), which may degrade the treatment accuracy, particularly in highly precise and conformal techniques. This study aims to quantitatively assess the dosimetric impact of systematic and random GPEs during the dynamic SPArc delivery.<i>Method.</i>Twolve clinical cases, including brain stereotactic radiosurgery (SRS), lung, liver, and prostate cancer, were evaluated under simulated systematic and random GPEs with ±0.5°, ±.1°, ±1.5°, ±2° per 180° arc scenarios. SPArc plans were generated using a published algorithm implemented in RayStation through scripting and simulated the treatment via a validated IBA DynamicARC® system model. Virtual machine logfiles reconstructed control points and spot positions, incorporating GPEs through modified gantry velocities. Target coverage (D98) and gamma passing ratios (GPR) under 3%/3 mm and 2%/2 mm criteria were analyzed.<i>Results.</i>Under the 3%/3 mm criterion, GPR exceeded 98% for all cases; liver, lung, and prostate plans remained >99% across GPE magnitudes, with lung, liver, and prostate plans exceeding 99.4% even at a 2° systematic error. Brain SRS cases were most sensitive; under a 2%/2 mm criterion, the GPR dropped to 89.52 ± 4.74% for a 2° systematic error. In contrast, lung, liver, and prostate plans maintained GPRs above 93% under the same conditions. The impact of random GPEs was minimal, with all plans achieving GPRs greater than 96% even for the strictest 2%/2 mm criterion.<i>Significance.</i>This study provides the first comprehensive quantitative evaluation of dose perturbation caused by GPE in the dynamic SPArc treatment delivery. Systematic GPEs introduce more significant dose deviations compared to random errors, with the impact varying by cases, disease sites and error magnitude. These findings could guide machine quality assurance protocols and site-specific error tolerance thresholds for future clinical implementation.

Physics in Medicine and BiologyVol. 70(17)
Union Hospital (HK), Wayne State University (US), Beaumont Hospital, Royal Oak (US), Beaumont Hospital, Troy (US), Huazhong University of Science and Technology (CN), Stanford University (US)
Openalex Percentile: Top 69%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.