CuO nanoparticle-enhanced electrochemical sensing of bromfenac in rabbit aqueous humor: a green analytical approach

Electrochemical techniques, particularly voltammetry, have emerged as powerful tools for the quantitative analysis of various analytes, including pharmaceuticals. Bromfenac (BRM), a widely used anti-inflammatory eye drop, is known for its rapid penetration into ocular tissues. Herein, a novel carbon paste electrode modified with synthesized copper oxide nanoparticles (CuO-NPs) is employed to enhance the electrochemical response for BRM. Differential pulse voltammetry (DPV) is utilized to measure the oxidation current of BRM in a Britton-Robinson buffer solution (pH 4.0) within a potential range of 0.0 V-1.5 V at a scan rate of 0.0125 V s<sup>-1</sup>. The developed method is validated according to ICH guidelines, and it exhibited excellent linearity in two concentration ranges: 2.00 × 10<sup>-7</sup> M-2.00 × 10<sup>-6</sup> M and 2.00 × 10<sup>-6</sup> M-2.00 × 10<sup>-5</sup> M. The accuracy of the method is found to be between 99.05% and 102.00%. The proposed voltammetric method is successfully applied for the determination of BRM in bulk powder, with a limit of detection (LOD) of 1.36 × 10<sup>-7</sup> M and a limit of quantification (LOQ) of 1.4 × 10<sup>-7</sup> M. This method is further extended to the quantification of BRM in a pharmaceutical ophthalmic solution and rabbit aqueous humor. To ensure the accuracy and precision of the developed method, the results of pharmaceutical formulation analysis are compared to a reported HPLC method, which showed no significant difference. Thus, this research demonstrates the potential of the developed electrochemical sensor for the accurate and reliable quantification of BRM in biological samples, providing valuable insights into its pharmacokinetics and pharmacodynamics. Eco-Scale (AES) and the Green Analytical Procedure Index (GAPI) tool are used to assess the greenness of this method, and this method proved to be environmentally friendly.

Authors

Institutions

Publication Details

Journal
Analytical Methods
Published
2025-01-01
DOI
https://doi.org/10.1039/d5ay01107h
Primary Topic
Electrochemical sensors and biosensors
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

CuO nanoparticle-enhanced electrochemical sensing of bromfenac in rabbit aqueous humor: a green analytical approach

Amr M. Mahmoud, Nermine S. Ghoniem, Loubna M. Elsharkawy, Samah S. Abbas et al.
Analytical Methods
Electrochemical sensors and biosensors
article

CuO nanoparticle-enhanced electrochemical sensing of bromfenac in rabbit aqueous humor: a green analytical approach

Amr M. Mahmoud, Nermine S. Ghoniem, Loubna M. Elsharkawy, Samah S. Abbas, Mariam Gamaleldin
article en

Abstract

Electrochemical techniques, particularly voltammetry, have emerged as powerful tools for the quantitative analysis of various analytes, including pharmaceuticals. Bromfenac (BRM), a widely used anti-inflammatory eye drop, is known for its rapid penetration into ocular tissues. Herein, a novel carbon paste electrode modified with synthesized copper oxide nanoparticles (CuO-NPs) is employed to enhance the electrochemical response for BRM. Differential pulse voltammetry (DPV) is utilized to measure the oxidation current of BRM in a Britton-Robinson buffer solution (pH 4.0) within a potential range of 0.0 V-1.5 V at a scan rate of 0.0125 V s<sup>-1</sup>. The developed method is validated according to ICH guidelines, and it exhibited excellent linearity in two concentration ranges: 2.00 × 10<sup>-7</sup> M-2.00 × 10<sup>-6</sup> M and 2.00 × 10<sup>-6</sup> M-2.00 × 10<sup>-5</sup> M. The accuracy of the method is found to be between 99.05% and 102.00%. The proposed voltammetric method is successfully applied for the determination of BRM in bulk powder, with a limit of detection (LOD) of 1.36 × 10<sup>-7</sup> M and a limit of quantification (LOQ) of 1.4 × 10<sup>-7</sup> M. This method is further extended to the quantification of BRM in a pharmaceutical ophthalmic solution and rabbit aqueous humor. To ensure the accuracy and precision of the developed method, the results of pharmaceutical formulation analysis are compared to a reported HPLC method, which showed no significant difference. Thus, this research demonstrates the potential of the developed electrochemical sensor for the accurate and reliable quantification of BRM in biological samples, providing valuable insights into its pharmacokinetics and pharmacodynamics. Eco-Scale (AES) and the Green Analytical Procedure Index (GAPI) tool are used to assess the greenness of this method, and this method proved to be environmentally friendly.

Analytical MethodsVol. 17(34)
Cairo University (EG), Future University in Egypt (EG), Nile University (EG), Egyptian Russian University (EG)
Faculty of Pharmacy, Cairo University
Clean water and sanitation
Openalex Percentile: Top 81%
Electrochemical sensors and biosensors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.