Experimental investigation of a neutron energy spectrometer in low-energy range using a position-sensitive proportional counter
Abstract This study investigated a novel spectrometer to measure the neutron energy spectrum in the epi-thermal range (0.5 eV to 10 keV) for Boron Neutron Capture Therapy (BNCT). A 3 He Position-Sensitive Proportional Counter (PSPC) was utilized to measure the depth-dependent distribution of neutron interactions within the detector (depth distribution). The energy spectrum was reconstructed using a spectrum unfolding technique based on Bayes' theorem. To demonstrate this approach, we conducted validation experiments using a custom-designed measurement system. A multi-parameter MCA was employed to detect two-dimensional reaction-depth distributions, from which the neutron spectrum was estimated. The results showed good agreement with the reference spectrum up to 1 keV, although side-neutron contributions were found to impact the estimation accuracy. This study serves as an important step to realize the feasibility of PSPC-based systems for low-energy neutron spectrometry and provides a foundation for further optimization to enhance estimation performance.
Authors
- Isao Murata (ORCID: https://orcid.org/0000-0003-2759-5561)
- Yuta Osawa
- Sachie Kusaka (ORCID: https://orcid.org/0009-0008-8141-7778)
- Fuminobu Sato (ORCID: https://orcid.org/0000-0001-9582-4915)
- Shingo Tamaki (ORCID: https://orcid.org/0000-0003-2159-1316)
- Y. Fujiwara (ORCID: https://orcid.org/0009-0008-5594-5972)
Institutions
- The University of Osaka (JP)
Publication Details
- Journal
- Journal of Instrumentation
- Published
- 2025-08-01
- DOI
- https://doi.org/10.1088/1748-0221/20/08/p08018
- Primary Topic
- Nuclear Physics and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00