Experimental investigation of a neutron energy spectrometer in low-energy range using a position-sensitive proportional counter

Abstract This study investigated a novel spectrometer to measure the neutron energy spectrum in the epi-thermal range (0.5 eV to 10 keV) for Boron Neutron Capture Therapy (BNCT). A 3 He Position-Sensitive Proportional Counter (PSPC) was utilized to measure the depth-dependent distribution of neutron interactions within the detector (depth distribution). The energy spectrum was reconstructed using a spectrum unfolding technique based on Bayes' theorem. To demonstrate this approach, we conducted validation experiments using a custom-designed measurement system. A multi-parameter MCA was employed to detect two-dimensional reaction-depth distributions, from which the neutron spectrum was estimated. The results showed good agreement with the reference spectrum up to 1 keV, although side-neutron contributions were found to impact the estimation accuracy. This study serves as an important step to realize the feasibility of PSPC-based systems for low-energy neutron spectrometry and provides a foundation for further optimization to enhance estimation performance.

Authors

Institutions

Publication Details

Journal
Journal of Instrumentation
Published
2025-08-01
DOI
https://doi.org/10.1088/1748-0221/20/08/p08018
Primary Topic
Nuclear Physics and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Experimental investigation of a neutron energy spectrometer in low-energy range using a position-sensitive proportional counter

Isao Murata, Yuta Osawa, Sachie Kusaka, Fuminobu Sato et al.
Journal of Instrumentation
Nuclear Physics and Applications
article

Experimental investigation of a neutron energy spectrometer in low-energy range using a position-sensitive proportional counter

Isao Murata, Yuta Osawa, Sachie Kusaka, Fuminobu Sato, Shingo Tamaki, Y. Fujiwara
article en

Abstract

Abstract This study investigated a novel spectrometer to measure the neutron energy spectrum in the epi-thermal range (0.5 eV to 10 keV) for Boron Neutron Capture Therapy (BNCT). A 3 He Position-Sensitive Proportional Counter (PSPC) was utilized to measure the depth-dependent distribution of neutron interactions within the detector (depth distribution). The energy spectrum was reconstructed using a spectrum unfolding technique based on Bayes' theorem. To demonstrate this approach, we conducted validation experiments using a custom-designed measurement system. A multi-parameter MCA was employed to detect two-dimensional reaction-depth distributions, from which the neutron spectrum was estimated. The results showed good agreement with the reference spectrum up to 1 keV, although side-neutron contributions were found to impact the estimation accuracy. This study serves as an important step to realize the feasibility of PSPC-based systems for low-energy neutron spectrometry and provides a foundation for further optimization to enhance estimation performance.

Journal of InstrumentationVol. 20(08)
The University of Osaka (JP)
Affordable and clean energy
Openalex Percentile: Top 72%
Nuclear Physics and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.