Rare Posterior Dislocation of Mobi-C Cervical Disc: A Case Report and Surgical Resolution

BACKGROUND Cervical disc prosthesis (CDP) is a well-established treatment for cervical degenerative disc disease. It provides motion preservation and decreases adjacent segment degeneration associated with ACDF. It is highly effective, with a low failure rate, but rare complications, including prosthesis migration, can occur. The posterior migration of a CDP is extremely rare, and only a few cases have been described in the literature. This report describes a unique case of posterior intraprosthetic dislocation of a Mobi-C prosthesis and adds valuable knowledge about this rare complication. CASE REPORT A 50-year-old female non-smoker underwent C6-C7 discectomy with implantation of the Mobi-C prosthesis for cervical disc herniation causing radicular pain and paresthesia. The initial postoperative results were good, with remission of neurological symptoms and no complications. However, after 2.5 years, she presented with neck pain and numbness of the right upper limb after routine physical activity. Imaging studies, including computed tomography (CT), revealed a posterior migration of the polyethylene insert with compression of the spinal cord. Revision surgery was undertaken comprising removal of the dislocated prosthesis and conversion to ACDF using the Anchor C system. Complete symptom resolution was achieved, and implants were properly aligned without complication in this procedure. CONCLUSIONS Posterior intraprosthetic dislocation of the Mobi-C prosthesis is an extremely rare complication. This case report with long-term follow-up shows ACDF is an effective method of salvage. Further investigation is necessary for understanding failure mechanisms, and development of CDP design is needed to avoid such complications.

Authors

Institutions

Publication Details

Journal
American Journal of Case Reports
Published
2025-08-10
DOI
https://doi.org/10.12659/ajcr.947721
Primary Topic
Spine and Intervertebral Disc Pathology
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Rare Posterior Dislocation of Mobi-C Cervical Disc: A Case Report and Surgical Resolution

Jurij Kseniuk, Artur Stolarczyk, Krzysztof Modzelewski, Atanas G. Atanasov et al.
American Journal of Case Reports
Spine and Intervertebral Disc Pathology
article

Rare Posterior Dislocation of Mobi-C Cervical Disc: A Case Report and Surgical Resolution

Jurij Kseniuk, Artur Stolarczyk, Krzysztof Modzelewski, Atanas G. Atanasov, Marcin Łapiński, Olga Adamska, Monika Wyszyńska
article en

Abstract

BACKGROUND Cervical disc prosthesis (CDP) is a well-established treatment for cervical degenerative disc disease. It provides motion preservation and decreases adjacent segment degeneration associated with ACDF. It is highly effective, with a low failure rate, but rare complications, including prosthesis migration, can occur. The posterior migration of a CDP is extremely rare, and only a few cases have been described in the literature. This report describes a unique case of posterior intraprosthetic dislocation of a Mobi-C prosthesis and adds valuable knowledge about this rare complication. CASE REPORT A 50-year-old female non-smoker underwent C6-C7 discectomy with implantation of the Mobi-C prosthesis for cervical disc herniation causing radicular pain and paresthesia. The initial postoperative results were good, with remission of neurological symptoms and no complications. However, after 2.5 years, she presented with neck pain and numbness of the right upper limb after routine physical activity. Imaging studies, including computed tomography (CT), revealed a posterior migration of the polyethylene insert with compression of the spinal cord. Revision surgery was undertaken comprising removal of the dislocated prosthesis and conversion to ACDF using the Anchor C system. Complete symptom resolution was achieved, and implants were properly aligned without complication in this procedure. CONCLUSIONS Posterior intraprosthetic dislocation of the Mobi-C prosthesis is an extremely rare complication. This case report with long-term follow-up shows ACDF is an effective method of salvage. Further investigation is necessary for understanding failure mechanisms, and development of CDP design is needed to avoid such complications.

American Journal of Case ReportsVol. 26
Medical University of Warsaw (PL), Cardinal Stefan Wyszynski University in Warsaw (PL), Medical University of Vienna (AT)
Gender equality
Openalex Percentile: Top 70%
Spine and Intervertebral Disc Pathology
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.