Magnetocardiography System With Self-Shielding Coils for High-Quality Cardiac Magnetic Signal Acquisition

To establish a near-zero magnetic environment for accurate magnetocardiography (MCG) measurements, a magnetically shielded cylinder (MSC) is commonly used in combination with active magnetic compensation coils. However, the coupling effect between the field from internal coils and the magnetic shielding layer can distort the distribution of ideal magnetic field and degrade the quality of cardiac magnetic signals. To address this issue, this article proposes a design approach for self-shielding coils featuring curved structures that integrate the target-field method with the pigeon-inspired optimization algorithm. The results indicate that the optimized coils maintain nonuniformity error below 1% in the internal region with a side length of 0.2 m and attain an attenuation error below 5% in the external area. This means that the coils are not easily affected by ferromagnetic materials and are suitable for MSCs of different sizes. An optically pumped magnetometer-based magnetocardiography (OPM-MCG) system, using two sets of single-axis curved coils, is used to compensate for the background field interference, resulting in an improved signal-to-noise ratio (SNR) for cardiac magnetic signals. Within the frequency range of 1–40 Hz, the average sensitivity of the OPM-MCG compensation system improves from 810.59 to 97.63 <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\text{fT/Hz}^{1/2}$ </tex-math></inline-formula>, and it achieves 9.19 dB of noise suppression. Based on the enhanced MCG system, the collected MCG signals reveal significant SNR gains of 6.44 and 5.89 dB in the R-wave and T-wave segments, which lays a foundation for further exploration of cardiac electrophysiological information and assisted cardiac treatment.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Instrumentation and Measurement
Published
2025-01-01
DOI
https://doi.org/10.1109/tim.2025.3597720
Citations
1
Primary Topic
Atomic and Subatomic Physics Research
Type
article
Field-Weighted Citation Impact
1.25

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Magnetocardiography System With Self-Shielding Coils for High-Quality Cardiac Magnetic Signal Acquisition

Xu Zhang, Ziyang Shi, Jianzhi Yang, Minxia Shi et al.
1 citations
IEEE Transactions on Instrumentation and Measurement
Atomic and Subatomic Physics Research
1.25
article

Magnetocardiography System With Self-Shielding Coils for High-Quality Cardiac Magnetic Signal Acquisition

Xu Zhang, Ziyang Shi, Jianzhi Yang, Minxia Shi, Kangqi Tian, Wenpeng Chen, Shiqiang Zheng, Ao Zhang, Leran Zhang
article en
1 citations

Abstract

To establish a near-zero magnetic environment for accurate magnetocardiography (MCG) measurements, a magnetically shielded cylinder (MSC) is commonly used in combination with active magnetic compensation coils. However, the coupling effect between the field from internal coils and the magnetic shielding layer can distort the distribution of ideal magnetic field and degrade the quality of cardiac magnetic signals. To address this issue, this article proposes a design approach for self-shielding coils featuring curved structures that integrate the target-field method with the pigeon-inspired optimization algorithm. The results indicate that the optimized coils maintain nonuniformity error below 1% in the internal region with a side length of 0.2 m and attain an attenuation error below 5% in the external area. This means that the coils are not easily affected by ferromagnetic materials and are suitable for MSCs of different sizes. An optically pumped magnetometer-based magnetocardiography (OPM-MCG) system, using two sets of single-axis curved coils, is used to compensate for the background field interference, resulting in an improved signal-to-noise ratio (SNR) for cardiac magnetic signals. Within the frequency range of 1–40 Hz, the average sensitivity of the OPM-MCG compensation system improves from 810.59 to 97.63 <inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\text{fT/Hz}^{1/2}$ </tex-math></inline-formula>, and it achieves 9.19 dB of noise suppression. Based on the enhanced MCG system, the collected MCG signals reveal significant SNR gains of 6.44 and 5.89 dB in the R-wave and T-wave segments, which lays a foundation for further exploration of cardiac electrophysiological information and assisted cardiac treatment.

IEEE Transactions on Instrumentation and MeasurementVol. 74
Beihang University (CN)
National Natural Science Foundation of China, National Postdoctoral Program for Innovative Talents
Affordable and clean energy
Openalex Percentile: Top 18%
Atomic and Subatomic Physics Research
1.25
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.