Correlation Analysis of Suture Anchor Pull-Out Strength with Cortical Bone Thickness and Cancellous Bone Density on a Finite Element Model
This study aimed to assess, using finite element analysis (FEA), the mechanical effects of cortical bone thickness and cancellous bone density on the pull-out strength of suture anchors. A PEEK anchor was modeled and embedded in synthetic bone blocks with cortical thicknesses ranging from 1 to 5 mm and cancellous densities of 10 PCF, 20 PCF, and 30 PCF. Axial tensile loading simulations were conducted for all combinations, and selected cases were validated through experimental pull-out tests using commercial synthetic bone, demonstrating agreement within ±6%. Both cortical thickness and cancellous density were found to enhance pull-out resistance, though the magnitude and pattern varied with density. At 10 PCF, pull-out strength increased linearly with cortical thickness. At 20 PCF, substantial gains were observed between 2 and 4 mm, followed by a plateau. At 30 PCF, most of the increase was confined between 2 and 3 mm, with minimal improvement thereafter. These findings suggest that fixation strategies should be adapted on the basis of bone quality and provide biomechanical insights to inform patient-specific implant design and surgical planning.
Authors
- J. HyeOk
- Jae Hyun Woo
- Jung Ho Kim (ORCID: https://orcid.org/0000-0003-4931-3553)
- Sung Min Kim (ORCID: https://orcid.org/0000-0001-6457-989X)
Institutions
- Dongguk University (KR)
Publication Details
- Journal
- Bioengineering
- Published
- 2025-08-11
- DOI
- https://doi.org/10.3390/bioengineering12080863
- Citations
- 3
- Primary Topic
- Dental Implant Techniques and Outcomes
- Type
- article
- Field-Weighted Citation Impact
- 5.37
Funders
- Ministry of Food and Drug Safety