<b>Background/Objectives</b>: Autorefractometers are valuable tools in clinical practice, but their accuracy is often questioned, especially in the pediatric population. This study aimed to compare refraction data from open-field and close-field autorefractometers and subjective refraction without using cycloplegia. <b>Methods</b>: A total of 50 eyes of 50 participants (19 males and 31 females, 11.8 ± 1.56 years) were evaluated. In a single visit, objective refraction was performed using NVision-K 5001 (open-field) and Visionix VX120 (close-field) autorefractometers, and subjective refraction using the fogging technique. Differences between procedures were assessed for sphere, spherical equivalent, and cylindrical vectors J<sub>0</sub> and J<sub>45</sub> using the Friedman test, followed by the post hoc Wilcoxon test as needed. <b>Results</b>: Significant differences were found in the sphere between the three procedures (all <i>p</i> ≤ 0.032). For the spherical equivalent, the Visionix VX120 differed significantly with the other two techniques (both <i>p</i> < 0.001), whereas no significant differences were found between NVision-K 5001 and subjective refraction (<i>p</i> = 0.193). Finally, no significant differences were observed for J<sub>0</sub> and J<sub>45</sub> vectors among the procedures (both <i>p</i> ≥ 0.166). <b>Conclusions</b>: There are certain discrepancies between autorefractometers and the subjective assessment of refractive error, most evident in measurements taken with the close-field device, possibly due to greater accommodative stimulation. However, in contexts such as visual screening or as a preliminary guide in the clinic, the values obtained by autorefractometry can provide useful information.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2025-08-11
DOI
https://doi.org/10.3390/jcm14165680
Primary Topic
Optical measurement and interference techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparing Close-Field and Open-Field Autorefractometry and Subjective Refraction

Hugo Pena‐Verdeal, Maria Jesus Giraldez, Eva Yebra‐Pimentel, Noelia Nores-Palmas et al.
Journal of Clinical Medicine
Optical measurement and interference techniques
article

Comparing Close-Field and Open-Field Autorefractometry and Subjective Refraction

Hugo Pena‐Verdeal, Maria Jesus Giraldez, Eva Yebra‐Pimentel, Noelia Nores-Palmas, Verónica Noya-Padin, Belén Sabucedo-Villamarin
article en

Abstract

<b>Background/Objectives</b>: Autorefractometers are valuable tools in clinical practice, but their accuracy is often questioned, especially in the pediatric population. This study aimed to compare refraction data from open-field and close-field autorefractometers and subjective refraction without using cycloplegia. <b>Methods</b>: A total of 50 eyes of 50 participants (19 males and 31 females, 11.8 ± 1.56 years) were evaluated. In a single visit, objective refraction was performed using NVision-K 5001 (open-field) and Visionix VX120 (close-field) autorefractometers, and subjective refraction using the fogging technique. Differences between procedures were assessed for sphere, spherical equivalent, and cylindrical vectors J<sub>0</sub> and J<sub>45</sub> using the Friedman test, followed by the post hoc Wilcoxon test as needed. <b>Results</b>: Significant differences were found in the sphere between the three procedures (all <i>p</i> ≤ 0.032). For the spherical equivalent, the Visionix VX120 differed significantly with the other two techniques (both <i>p</i> < 0.001), whereas no significant differences were found between NVision-K 5001 and subjective refraction (<i>p</i> = 0.193). Finally, no significant differences were observed for J<sub>0</sub> and J<sub>45</sub> vectors among the procedures (both <i>p</i> ≥ 0.166). <b>Conclusions</b>: There are certain discrepancies between autorefractometers and the subjective assessment of refractive error, most evident in measurements taken with the close-field device, possibly due to greater accommodative stimulation. However, in contexts such as visual screening or as a preliminary guide in the clinic, the values obtained by autorefractometry can provide useful information.

Journal of Clinical MedicineVol. 14(16)
Universidade de Santiago de Compostela (ES), Instituto de Investigación Sanitaria de Santiago (ES)
Openalex Percentile: Top 80%
Optical measurement and interference techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.