Comparing Close-Field and Open-Field Autorefractometry and Subjective Refraction
<b>Background/Objectives</b>: Autorefractometers are valuable tools in clinical practice, but their accuracy is often questioned, especially in the pediatric population. This study aimed to compare refraction data from open-field and close-field autorefractometers and subjective refraction without using cycloplegia. <b>Methods</b>: A total of 50 eyes of 50 participants (19 males and 31 females, 11.8 ± 1.56 years) were evaluated. In a single visit, objective refraction was performed using NVision-K 5001 (open-field) and Visionix VX120 (close-field) autorefractometers, and subjective refraction using the fogging technique. Differences between procedures were assessed for sphere, spherical equivalent, and cylindrical vectors J<sub>0</sub> and J<sub>45</sub> using the Friedman test, followed by the post hoc Wilcoxon test as needed. <b>Results</b>: Significant differences were found in the sphere between the three procedures (all <i>p</i> ≤ 0.032). For the spherical equivalent, the Visionix VX120 differed significantly with the other two techniques (both <i>p</i> < 0.001), whereas no significant differences were found between NVision-K 5001 and subjective refraction (<i>p</i> = 0.193). Finally, no significant differences were observed for J<sub>0</sub> and J<sub>45</sub> vectors among the procedures (both <i>p</i> ≥ 0.166). <b>Conclusions</b>: There are certain discrepancies between autorefractometers and the subjective assessment of refractive error, most evident in measurements taken with the close-field device, possibly due to greater accommodative stimulation. However, in contexts such as visual screening or as a preliminary guide in the clinic, the values obtained by autorefractometry can provide useful information.
Authors
- Hugo Pena‐Verdeal (ORCID: https://orcid.org/0000-0001-5103-1374)
- Maria Jesus Giraldez (ORCID: https://orcid.org/0000-0001-9595-7486)
- Eva Yebra‐Pimentel (ORCID: https://orcid.org/0000-0002-4012-4636)
- Noelia Nores-Palmas (ORCID: https://orcid.org/0000-0001-8317-8039)
- Verónica Noya-Padin (ORCID: https://orcid.org/0000-0002-2815-3653)
- Belén Sabucedo-Villamarin (ORCID: https://orcid.org/0000-0001-5032-620X)
Institutions
- Universidade de Santiago de Compostela (ES)
- Instituto de Investigación Sanitaria de Santiago (ES)
Publication Details
- Journal
- Journal of Clinical Medicine
- Published
- 2025-08-11
- DOI
- https://doi.org/10.3390/jcm14165680
- Primary Topic
- Optical measurement and interference techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00