Abstract In the solar reflective band, the on-board calibration method based on a solar diffuser (SD) is the mainstream calibration method. However, the calibration accuracy is limited by the measurement accuracy of the SD bidirectional reflectance distribution function (BRDF) in the laboratory. The limitations of the light sources and detectors hinder accurate BRDF measurements in the short-wave infrared band, particularly beyond 1700 nm. In this study, the angle integration-band crossing transfer (AIBCT) method is proposed for SD BRDF measurement to avoid the influence of the Fourier spectrometer’s signal processing and detector non-linearity in the absolute measurement. The method utilized the relationship between the BRDF and directional-hemispherical reflectance, and calculated the spectral distribution of the SD BRDF at the standard in-plane geometry with an 0° incidence zenith angle and a 45° reflection zenith angle (0:45) by integration. The absolute measurement result of the 0:45 SD BRDF at 1150 nm was used as the standard for transfer to the short-wave infrared band. The SD BRDF at 0:45 was generalized to other geometries through relative measurements based on the reciprocity theorem. The mean relative error of the results of the AIBCT and absolute measurement methods was less than 0.3% in the range of 1100 nm to 1600 nm. Analysis of the results showed that the AIBCT method’s measurement uncertainty (coverage factor k = 2) of the 0:45 SD BRDF is better than 0.84% at 1100 nm to 2300 nm and 1.05% at 2300 nm to 2500 nm. Furthermore, the AIBCT method was found suitable for the ultraviolet band.

Authors

Institutions

Publication Details

Journal
Metrologia
Published
2025-08-01
DOI
https://doi.org/10.1088/1681-7575/adfa9c
Primary Topic
Calibration and Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Measurement method of the solar diffuser bidirectional reflectance distribution function in the short-wave infrared band

Zhiyuan Zhang, Xiaolong Si, Xiaobing Zheng, Xin Li et al.
Metrologia
Calibration and Measurement Techniques
article

Measurement method of the solar diffuser bidirectional reflectance distribution function in the short-wave infrared band

Zhiyuan Zhang, Xiaolong Si, Xiaobing Zheng, Xin Li, Wenxin Huang, Hongyao Chen
article en

Abstract

Abstract In the solar reflective band, the on-board calibration method based on a solar diffuser (SD) is the mainstream calibration method. However, the calibration accuracy is limited by the measurement accuracy of the SD bidirectional reflectance distribution function (BRDF) in the laboratory. The limitations of the light sources and detectors hinder accurate BRDF measurements in the short-wave infrared band, particularly beyond 1700 nm. In this study, the angle integration-band crossing transfer (AIBCT) method is proposed for SD BRDF measurement to avoid the influence of the Fourier spectrometer’s signal processing and detector non-linearity in the absolute measurement. The method utilized the relationship between the BRDF and directional-hemispherical reflectance, and calculated the spectral distribution of the SD BRDF at the standard in-plane geometry with an 0° incidence zenith angle and a 45° reflection zenith angle (0:45) by integration. The absolute measurement result of the 0:45 SD BRDF at 1150 nm was used as the standard for transfer to the short-wave infrared band. The SD BRDF at 0:45 was generalized to other geometries through relative measurements based on the reciprocity theorem. The mean relative error of the results of the AIBCT and absolute measurement methods was less than 0.3% in the range of 1100 nm to 1600 nm. Analysis of the results showed that the AIBCT method’s measurement uncertainty (coverage factor k = 2) of the 0:45 SD BRDF is better than 0.84% at 1100 nm to 2300 nm and 1.05% at 2300 nm to 2500 nm. Furthermore, the AIBCT method was found suitable for the ultraviolet band.

MetrologiaVol. 62(4)
University of Science and Technology of China (CN), Hefei Institutes of Physical Science (CN)
Openalex Percentile: Top 71%
Calibration and Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.