The biocompatibility of a novel polyether ether ketone (PEEK) stoma device with human epidermal keratinocytes

Surgical stomas are essential interventions for many medical conditions, however, can create physical complications, such as peristomal skin irritation. Medical stoma devices (MSD) are a potential treatment to mitigate such complications. Polyether ether ketone (PEEK) is a promising material for implantable MSD due to its established biocompatibility within orthodontic and orthopaedic applications. However, its cytocompatibility with human epidermal keratinocytes has not been evaluated according to ISO 10993-5:2009 guidelines. This study aimed to assess the biocompatibility of a novel PEEK MSD with human keratinocytes (HaCaT cells). Cells were cultured on discs of PEEK, surface-modified PEEK (m-PEEK), and polylactic acid (PLA). Surface topography was investigated via SEM to assess surface roughness (S<sub>a</sub>, S<sub>z</sub>), and water contact angle (WCA). m-PEEK demonstrated increased S<sub>a</sub> and WCA compared to PLA and unmodified PEEK. Cell proliferation and viability were evaluated using CyQUANT™ and AlamarBlue™ assays and no significant differences were observed among PLA, PEEK, and m-PEEK. Cell adhesion was assessed using an adhesion assay, with m-PEEK demonstrating significantly higher cell adhesion than PLA (p < 0.05), with cell attachment confirmed via SEM imaging. Cytokine analysis of supernatants using Luminex Immunoassay revealed two (IL-1α and IL-6) of six cytokines outlined in the ISO 10993-20:2006 guidelines were elevated in the presence of PEEK at 72 h. These findings suggest that PEEK is non-cytotoxic and biocompatible with human keratinocytes. Further studies are warranted to assess PEEK's compatibility with colonic cells, 3D skin models, and in vivo systems (including for chronic inflammatory responses) for MSD applications.

Authors

Institutions

Publication Details

Journal
Biomaterials Advances
Published
2025-08-10
DOI
https://doi.org/10.1016/j.bioadv.2025.214459
Primary Topic
Wound Healing and Treatments
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The biocompatibility of a novel polyether ether ketone (PEEK) stoma device with human epidermal keratinocytes

Paulomi Polly Burey, Eliza Whiteside, Anna Chruścik, Andreas Marc Palmer et al.
Biomaterials Advances
Wound Healing and Treatments
article

The biocompatibility of a novel polyether ether ketone (PEEK) stoma device with human epidermal keratinocytes

Paulomi Polly Burey, Eliza Whiteside, Anna Chruścik, Andreas Marc Palmer, Racheal Wadlow, Mathilde Maybery, Nikita Walz
article en

Abstract

Surgical stomas are essential interventions for many medical conditions, however, can create physical complications, such as peristomal skin irritation. Medical stoma devices (MSD) are a potential treatment to mitigate such complications. Polyether ether ketone (PEEK) is a promising material for implantable MSD due to its established biocompatibility within orthodontic and orthopaedic applications. However, its cytocompatibility with human epidermal keratinocytes has not been evaluated according to ISO 10993-5:2009 guidelines. This study aimed to assess the biocompatibility of a novel PEEK MSD with human keratinocytes (HaCaT cells). Cells were cultured on discs of PEEK, surface-modified PEEK (m-PEEK), and polylactic acid (PLA). Surface topography was investigated via SEM to assess surface roughness (S<sub>a</sub>, S<sub>z</sub>), and water contact angle (WCA). m-PEEK demonstrated increased S<sub>a</sub> and WCA compared to PLA and unmodified PEEK. Cell proliferation and viability were evaluated using CyQUANT™ and AlamarBlue™ assays and no significant differences were observed among PLA, PEEK, and m-PEEK. Cell adhesion was assessed using an adhesion assay, with m-PEEK demonstrating significantly higher cell adhesion than PLA (p < 0.05), with cell attachment confirmed via SEM imaging. Cytokine analysis of supernatants using Luminex Immunoassay revealed two (IL-1α and IL-6) of six cytokines outlined in the ISO 10993-20:2006 guidelines were elevated in the presence of PEEK at 72 h. These findings suggest that PEEK is non-cytotoxic and biocompatible with human keratinocytes. Further studies are warranted to assess PEEK's compatibility with colonic cells, 3D skin models, and in vivo systems (including for chronic inflammatory responses) for MSD applications.

Biomaterials AdvancesVol. 178
University of Southern Queensland (AU)
University of Southern Queensland
Zero hunger
Openalex Percentile: Top 74%
Wound Healing and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.