Development of a versatile image quality test phantom for single photon emission computed tomography system

Single photon emission computed tomography (SPECT) is a vital imaging tool for visualizing functional and physiological processes of human body. Periodic quality assurance (QA) ensures imaging accuracy, but high-cost commercial phantoms limit accessibility in resource-constrained settings. To address this, a cost-effective and modular SPECT phantom was developed to evaluate imaging parameters in a single scan. The phantom, made from PMMA, features interchangeable modules designed to assess tomographic resolution, tomographic uniformity, tomographic contrast, and slice thickness. Its cylindrical structure accommodates radionuclide solutions to simulate imaging conditions. Validation was conducted using<sup>99m</sup>Tc and a dual-head SPECT/CT scanner (Discovery 670 DR, GE Healthcare, Chicago, USA) equipped with low-energy high-resolution (LEHR) collimators. Imaging data were analysed for pixel size accuracy, tomographic resolution, tomographic uniformity, slice thickness and tomographic contrast. The measured pixel size was 1.11 mm (<i>x</i>-axis) and 1.12 mm (<i>y</i>-axis), aligning closely with the system-calculated size of 1.10 mm. The tomographic resolution in air, assessed with the OSEM algorithm (2 iterations, 10 subsets) with post-reconstruction filtering using Butterworth filter (cutoff frequency: 0.48; order: 10), achieved a horizontal FWHM of 13.12 mm, superior to results from filtered back-projection (FBP). Tomographic uniformity assessments showed contrast variations within 3.6%, meeting IAEA standards. Cold rod and sphere contrast tests detected rods as small as 6.5 mm and spheres down to 9 mm diameter, while hot rod contrast test identified rods as small as 9 mm. Measurement results of our phantom was also found comparable with the measurements results of Jaszczak phantom. The test results indicate that our image quality test phantom is suitable for comprehensive QA of SPECT system. Its modular design offers additional advantage to add new test sections or modify currently available test sections without changing the core design of the phantom. Hence, it is a versatile cost-effective solution for the periodic QA in nuclear medicine.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2025-08-14
DOI
https://doi.org/10.1088/2057-1976/adfb82
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a versatile image quality test phantom for single photon emission computed tomography system

B.L. Malpani, Balvinder Kaur Sapra, Ankit Srivastava, Rajesh Kumar et al.
Biomedical Physics & Engineering Express
Medical Imaging Techniques and Applications
article

Development of a versatile image quality test phantom for single photon emission computed tomography system

B.L. Malpani, Balvinder Kaur Sapra, Ankit Srivastava, Rajesh Kumar, Sutapa Rakshit, Sunil Dutt Sharma
article en

Abstract

Single photon emission computed tomography (SPECT) is a vital imaging tool for visualizing functional and physiological processes of human body. Periodic quality assurance (QA) ensures imaging accuracy, but high-cost commercial phantoms limit accessibility in resource-constrained settings. To address this, a cost-effective and modular SPECT phantom was developed to evaluate imaging parameters in a single scan. The phantom, made from PMMA, features interchangeable modules designed to assess tomographic resolution, tomographic uniformity, tomographic contrast, and slice thickness. Its cylindrical structure accommodates radionuclide solutions to simulate imaging conditions. Validation was conducted using<sup>99m</sup>Tc and a dual-head SPECT/CT scanner (Discovery 670 DR, GE Healthcare, Chicago, USA) equipped with low-energy high-resolution (LEHR) collimators. Imaging data were analysed for pixel size accuracy, tomographic resolution, tomographic uniformity, slice thickness and tomographic contrast. The measured pixel size was 1.11 mm (<i>x</i>-axis) and 1.12 mm (<i>y</i>-axis), aligning closely with the system-calculated size of 1.10 mm. The tomographic resolution in air, assessed with the OSEM algorithm (2 iterations, 10 subsets) with post-reconstruction filtering using Butterworth filter (cutoff frequency: 0.48; order: 10), achieved a horizontal FWHM of 13.12 mm, superior to results from filtered back-projection (FBP). Tomographic uniformity assessments showed contrast variations within 3.6%, meeting IAEA standards. Cold rod and sphere contrast tests detected rods as small as 6.5 mm and spheres down to 9 mm diameter, while hot rod contrast test identified rods as small as 9 mm. Measurement results of our phantom was also found comparable with the measurements results of Jaszczak phantom. The test results indicate that our image quality test phantom is suitable for comprehensive QA of SPECT system. Its modular design offers additional advantage to add new test sections or modify currently available test sections without changing the core design of the phantom. Hence, it is a versatile cost-effective solution for the periodic QA in nuclear medicine.

Biomedical Physics & Engineering ExpressVol. 11(5)
Bhabha Atomic Research Centre (IN), Homi Bhabha National Institute (IN), Bhabha Atomic Research Center Hospital (IN)
Openalex Percentile: Top 71%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.