Comparative Performance of the CamPROBE Local Anaesthetic Transperineal Biopsy Device Versus an In-Line Device for Detection of Significant Prostate Cancer

<b>Background/Objective:</b> The CamPROBE device offers a simple, low-cost method to facilitate double-free-hand local anaesthetic transperineal prostate biopsies (LATPBx). Here we present data from prospective implementation of its use for first biopsy cancer detection. <b>Methods</b>: The outcomes of two centres who adopted the CamPROBE were compared to a retrospective series of biopsies using an in-line (single-free-hand) device. All biopsies were done by clinicians new to the device. Outcomes were the detection of any cancer and clinically significant prostate cancer (csPCa) defined as ≥Grade Group 2 and ≥Grade Group 3 (GG2/3) and composite ≥NICE Cambridge Prognostic Group 2 and 3 (CPG2/3), as well as sampling accuracy of MRI-defined lesions. <b>Results:</b> Device cohorts were well matched for pre-biopsy PSA, T stage, and MRI visibility in both centres. In centre 1 (100 CamPROBE vs. 97 in-line biopsies), there were no differences in detection of any cancer or csPCa: ≥GG2 60.0% vs. 56.7% (<i>p</i> = 0.64), ≥GG3 31% vs. 20.6% (<i>p</i> = 0.09), ≥CPG2 62.0% vs. 60.8% (<i>p</i> = 0.86), and ≥CPG3 (<i>p</i> = 0.55). There were also no differences between devices in target biopsy positivity: 67.4% vs. 63.5% (<i>p</i> = 0.59). Data from centre 2 (38 CamPROBE vs. 44 in-line) re-capitulated these findings. The MRI target detection rate of the CamPROBE (assessed in Centre 1) was not affected by prostate volume, lesion laterality, anatomical position, or lesion size. Limitations include modest sample sizes, lack of randomization, and patient-reported outcomes. <b>Conclusions</b>: These data demonstrate that the CamPROBE device is a highly effective method of performing prostate biopsies with excellent cancer detection rates and accuracy, supporting its wider dissemination and use.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2025-08-12
DOI
https://doi.org/10.3390/jcm14165702
Primary Topic
Prostate Cancer Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative Performance of the CamPROBE Local Anaesthetic Transperineal Biopsy Device Versus an In-Line Device for Detection of Significant Prostate Cancer

Vincent J. Gnanapragasam, Saiful Miah, Syed Kashif Ali Shah, Hannah Thorman et al.
Journal of Clinical Medicine
Prostate Cancer Diagnosis and Treatment
article

Comparative Performance of the CamPROBE Local Anaesthetic Transperineal Biopsy Device Versus an In-Line Device for Detection of Significant Prostate Cancer

Vincent J. Gnanapragasam, Saiful Miah, Syed Kashif Ali Shah, Hannah Thorman, Adham Ahmed, Kasra Saeb‐Parsy, Kieran Sandhu
article en

Abstract

<b>Background/Objective:</b> The CamPROBE device offers a simple, low-cost method to facilitate double-free-hand local anaesthetic transperineal prostate biopsies (LATPBx). Here we present data from prospective implementation of its use for first biopsy cancer detection. <b>Methods</b>: The outcomes of two centres who adopted the CamPROBE were compared to a retrospective series of biopsies using an in-line (single-free-hand) device. All biopsies were done by clinicians new to the device. Outcomes were the detection of any cancer and clinically significant prostate cancer (csPCa) defined as ≥Grade Group 2 and ≥Grade Group 3 (GG2/3) and composite ≥NICE Cambridge Prognostic Group 2 and 3 (CPG2/3), as well as sampling accuracy of MRI-defined lesions. <b>Results:</b> Device cohorts were well matched for pre-biopsy PSA, T stage, and MRI visibility in both centres. In centre 1 (100 CamPROBE vs. 97 in-line biopsies), there were no differences in detection of any cancer or csPCa: ≥GG2 60.0% vs. 56.7% (<i>p</i> = 0.64), ≥GG3 31% vs. 20.6% (<i>p</i> = 0.09), ≥CPG2 62.0% vs. 60.8% (<i>p</i> = 0.86), and ≥CPG3 (<i>p</i> = 0.55). There were also no differences between devices in target biopsy positivity: 67.4% vs. 63.5% (<i>p</i> = 0.59). Data from centre 2 (38 CamPROBE vs. 44 in-line) re-capitulated these findings. The MRI target detection rate of the CamPROBE (assessed in Centre 1) was not affected by prostate volume, lesion laterality, anatomical position, or lesion size. Limitations include modest sample sizes, lack of randomization, and patient-reported outcomes. <b>Conclusions</b>: These data demonstrate that the CamPROBE device is a highly effective method of performing prostate biopsies with excellent cancer detection rates and accuracy, supporting its wider dissemination and use.

Journal of Clinical MedicineVol. 14(16)
University of Cambridge (GB), Cambridge University Hospitals NHS Foundation Trust (GB), Bridge University (SS), Prostate Cancer Research (GB), West Suffolk Hospital (GB), Prostate Cancer UK (GB)
National Institute for Health and Care Research, Department of Health and Social Care, NIHR Cambridge Biomedical Research Centre
Openalex Percentile: Top 69%
Prostate Cancer Diagnosis and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.