Performance evaluation of the first modules of the of the gradient-phase interferometer within the ELT M1 Local Coherencer

An instrument for phasing out the segments of the primary mirror of the Extremely Large Telescope (ELT) is being developed by IDOM and UPC in collaboration with ESO. The sensing module includes a partially coherent interferometer with a prism introducing a continuous phase gradient between the interferometer arms, and a stepped glass arrangement to cope with a measurement range of ±250μm with a 2σ accuracy within ±300nm. The setup also allows direct independent measurements of relative tip and tilt. This paper shows the detailed calibration setup, the procedures performed in detail, and the results obtained. The calibration procedures show a 2σ average accuracy of 164nm across the full measurement range, and 16μrad and 28μrad for relative tip and tilt, respectively.

Authors

Institutions

Publication Details

Published
2025-08-08
DOI
https://doi.org/10.1117/12.3066441
Primary Topic
Photonic and Optical Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance evaluation of the first modules of the of the gradient-phase interferometer within the ELT M1 Local Coherencer

Noel Rodrigo, Martin Dimmler, Gaizka Murga, Maialen González et al.
Photonic and Optical Devices
article

Performance evaluation of the first modules of the of the gradient-phase interferometer within the ELT M1 Local Coherencer

Noel Rodrigo, Martin Dimmler, Gaizka Murga, Maialen González, Borja Vega, Santiago Royo, Philippe B. Gitton, Alexander Díaz, Andreas Förster, Juan Francisco Márquez, Sebastian P. Schmid, Afonso Teixeira, Samuel Lévêque, Pau Santos
article en

Abstract

An instrument for phasing out the segments of the primary mirror of the Extremely Large Telescope (ELT) is being developed by IDOM and UPC in collaboration with ESO. The sensing module includes a partially coherent interferometer with a prism introducing a continuous phase gradient between the interferometer arms, and a stepped glass arrangement to cope with a measurement range of ±250μm with a 2σ accuracy within ±300nm. The setup also allows direct independent measurements of relative tip and tilt. This paper shows the detailed calibration setup, the procedures performed in detail, and the results obtained. The calibration procedures show a 2σ average accuracy of 164nm across the full measurement range, and 16μrad and 28μrad for relative tip and tilt, respectively.

Idom (Spain) (ES), Universitat Politècnica de Catalunya (ES)
Openalex Percentile: Top 81%
Photonic and Optical Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Performance evaluation of the first modules of the of the gradient-phase interferometer within the ELT M1 Local Coherencer — Noel Rodrigo, Martin Dimmler, et al. · (2025) | TGRS Research Map | TGRS