Development and Characterisation of a Microneedle Sensor for Intrapartum Fetal Monitoring

This study presents the in vitro and preliminary ex vivo development of a novel microneedle-based pH sensor for continuous intrapartum fetal monitoring. The objective was to evaluate the feasibility of using microneedle sensors to monitor fetal pH during labour and to develop a proof-of-principle microneedle pH sensor that meets clinical requirements such as high sensitivity to small pH changes (0.05 units) within a relevant range (6.50-7.45), minimal tissue disruption, and a compact design suitable for transcervical placement on the fetal scalp (<40 mm diameter). Platinum microneedles were passivated with ArCare medical adhesive and coated with iridium oxide via electrodeposition. Sensitivity was tested in phosphate buffered saline (PBS) and artificial interstitial fluid (ISF), using both external Ag/AgCl and internal platinum pseudo-reference electrodes. In PBS, the sensor exhibited linear responses in increments of 0.05 pH units over the clinically relevant range (6.5-7.45), with slopes of -60.49 mV/pH (R<sup>2</sup> = 0.946, accuracy = 97.65%) and -63.2 mV/pH (R<sup>2</sup> = 0.910, accuracy = 93.70%) in the external and internal configurations, respectively. In ISF, a slope of -25.5 mV/pH (R<sup>2</sup> = 0.979) was obtained. Ex vivo testing on human skin confirmed successful microneedle penetration without visible iridium oxide transfer or tissue damage, as indicated by methylene blue staining. These findings support the potential for continuous minimally invasive fetal pH monitoring during labour, representing a significant step toward more objective and specific intrapartum assessment.

Authors

Institutions

Publication Details

Journal
Biosensors
Published
2025-08-08
DOI
https://doi.org/10.3390/bios15080517
Primary Topic
Advancements in Transdermal Drug Delivery
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and Characterisation of a Microneedle Sensor for Intrapartum Fetal Monitoring

Richard A. Greene, Paul Galvin, Conor O’Mahony, Sofia Rodrigues Teixeira et al.
Biosensors
Advancements in Transdermal Drug Delivery
article

Development and Characterisation of a Microneedle Sensor for Intrapartum Fetal Monitoring

Richard A. Greene, Paul Galvin, Conor O’Mahony, Sofia Rodrigues Teixeira, Fergus P. McCarthy, Rafael Sebastián, Jill Mitchell, John R. Higgins, Chinmay Thatte
article en

Abstract

This study presents the in vitro and preliminary ex vivo development of a novel microneedle-based pH sensor for continuous intrapartum fetal monitoring. The objective was to evaluate the feasibility of using microneedle sensors to monitor fetal pH during labour and to develop a proof-of-principle microneedle pH sensor that meets clinical requirements such as high sensitivity to small pH changes (0.05 units) within a relevant range (6.50-7.45), minimal tissue disruption, and a compact design suitable for transcervical placement on the fetal scalp (<40 mm diameter). Platinum microneedles were passivated with ArCare medical adhesive and coated with iridium oxide via electrodeposition. Sensitivity was tested in phosphate buffered saline (PBS) and artificial interstitial fluid (ISF), using both external Ag/AgCl and internal platinum pseudo-reference electrodes. In PBS, the sensor exhibited linear responses in increments of 0.05 pH units over the clinically relevant range (6.5-7.45), with slopes of -60.49 mV/pH (R<sup>2</sup> = 0.946, accuracy = 97.65%) and -63.2 mV/pH (R<sup>2</sup> = 0.910, accuracy = 93.70%) in the external and internal configurations, respectively. In ISF, a slope of -25.5 mV/pH (R<sup>2</sup> = 0.979) was obtained. Ex vivo testing on human skin confirmed successful microneedle penetration without visible iridium oxide transfer or tissue damage, as indicated by methylene blue staining. These findings support the potential for continuous minimally invasive fetal pH monitoring during labour, representing a significant step toward more objective and specific intrapartum assessment.

BiosensorsVol. 15(8)
National University of Ireland (IE), University College Cork (IE), Infant (IE), International Energy Research Centre (IE)
Enterprise Ireland, Higher Education Authority, European Regional Development Fund
Openalex Percentile: Top 85%
Advancements in Transdermal Drug Delivery
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.