Design and evaluation of multimodal concentric surface coils for 3T MR imaging

Motivation: In 3T MRI, achieving optimal B1 field efficiency and reducing specific absorption rate (SAR) are essential for enhancing image quality and patient safety. Goal(s): To develop a multimodal concentric surface coil that improves B1 field efficiency and reduces SAR compared to conventional surface coils. Approach: The novel coil design incorporates five electromagnetically coupled concentric resonators to form a multimodal resonator. Performance was evaluated through simulations and bench tests. Results: Both simulations and bench tests demonstrate that the concentric coil achieves superior B1 efficiency and lower SAR, underscoring its potential to improve image quality and safety in in vivo MRI applications. Impact: The proposed multimodal concentric coil demonstrates superior B1 field efficiency and reduced SAR compared to conventional surface coils. These improvements can significantly enhance image quality and patient safety in 3T MRI, paving the way for better clinical outcomes.

Authors

Institutions

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2025-09-16
DOI
https://doi.org/10.58530/2025/4246
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and evaluation of multimodal concentric surface coils for 3T MR imaging

Xiaoliang Zhang, Yunkun Zhao, Aditya Ashok Bhosale
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

Design and evaluation of multimodal concentric surface coils for 3T MR imaging

Xiaoliang Zhang, Yunkun Zhao, Aditya Ashok Bhosale
article en

Abstract

Motivation: In 3T MRI, achieving optimal B1 field efficiency and reducing specific absorption rate (SAR) are essential for enhancing image quality and patient safety. Goal(s): To develop a multimodal concentric surface coil that improves B1 field efficiency and reduces SAR compared to conventional surface coils. Approach: The novel coil design incorporates five electromagnetically coupled concentric resonators to form a multimodal resonator. Performance was evaluated through simulations and bench tests. Results: Both simulations and bench tests demonstrate that the concentric coil achieves superior B1 efficiency and lower SAR, underscoring its potential to improve image quality and safety in in vivo MRI applications. Impact: The proposed multimodal concentric coil demonstrates superior B1 field efficiency and reduced SAR compared to conventional surface coils. These improvements can significantly enhance image quality and patient safety in 3T MRI, paving the way for better clinical outcomes.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and ExhibitionVol. 33
University at Buffalo, State University of New York (US)
Openalex Percentile: Top 71%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.