The Usefulness of the Air Hole Phantom for Evaluating Z-resolution and Geometric Distortion in Digital Breast Tomosynthesis

Currently, the quality control of digital breast tomosynthesis (DBT) functionality in breast X-ray equipment is not adequately implemented. In particular, DBT involves reconstruction, so depth resolution and distortion need to be evaluated. In this study, we fabricated the air hole phantoms based on the Digital Breast Tomosynthesis Quality Control Manual Ver. 1.0 (hereinafter referred to as the DBT quality control manual) and examined their usefulness as phantoms for measuring Z-resolution and geometric distortion. Using the prototype phantoms, we measured Z-resolution and geometric distortion according to the DBT quality control manual. The results of the aluminum sphere phantoms and the air hole phantoms confirmed that the z-FWHM was within or below the reference value range in the technical evaluation report by the National Coordinating Centre for the Physics of Mammography. The geometric distortion was within the specified range in the DBT quality control manual (within percentage scaling error ±5%). Therefore, it was suggested that the air-hole phantom we developed is useful for measuring the Z-resolution and geometric distortion of DBT.

Authors

Institutions

Publication Details

Journal
Japanese Journal of Radiological Technology
Published
2025-01-01
DOI
https://doi.org/10.6009/jjrt.25-1559
Primary Topic
Digital Radiography and Breast Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Usefulness of the Air Hole Phantom for Evaluating Z-resolution and Geometric Distortion in Digital Breast Tomosynthesis

Tsuyoshi Yamada, Yasutaka Ichikawa, Yoya Tomita, Hajime Sakuma et al.
1 links
Japanese Journal of Radiological Technology
Digital Radiography and Breast Imaging
article

The Usefulness of the Air Hole Phantom for Evaluating Z-resolution and Geometric Distortion in Digital Breast Tomosynthesis

Tsuyoshi Yamada, Yasutaka Ichikawa, Yoya Tomita, Hajime Sakuma, Norio Naito, Norimitsu Shinohara
article en

Abstract

Currently, the quality control of digital breast tomosynthesis (DBT) functionality in breast X-ray equipment is not adequately implemented. In particular, DBT involves reconstruction, so depth resolution and distortion need to be evaluated. In this study, we fabricated the air hole phantoms based on the Digital Breast Tomosynthesis Quality Control Manual Ver. 1.0 (hereinafter referred to as the DBT quality control manual) and examined their usefulness as phantoms for measuring Z-resolution and geometric distortion. Using the prototype phantoms, we measured Z-resolution and geometric distortion according to the DBT quality control manual. The results of the aluminum sphere phantoms and the air hole phantoms confirmed that the z-FWHM was within or below the reference value range in the technical evaluation report by the National Coordinating Centre for the Physics of Mammography. The geometric distortion was within the specified range in the DBT quality control manual (within percentage scaling error ±5%). Therefore, it was suggested that the air-hole phantom we developed is useful for measuring the Z-resolution and geometric distortion of DBT.

Japanese Journal of Radiological TechnologyVol. 81(10)
Mie University (JP), Gifu University of Medical Science (JP), Mie University Hospital (JP)
Department of Biotechnology, Ministry of Science and Technology, India
Openalex Percentile: Top 69%
Digital Radiography and Breast Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.