Bicarbonate Buffer Dissolution Test Using the Floating Lid Method: Inter-Laboratory Reproducibility of pH Maintenance

Bicarbonate buffer (BCB) has been difficult to use in conventional dissolution tests because its pH rapidly increases as CO<sub>2</sub> escapes from the air-water interface. Recently, the floating lid method was introduced as a convenient method for using BCB in dissolution tests. This study aimed to confirm the inter-laboratory reproducibility of pH maintenance of BCB using the floating lid method for both paddle and flow-through cell (FTC) methods. Three pharmaceutical companies and 1 academic research institute participated in this study. A BCB solution (pH 6.5, 15 mM) was employed as the test solution. In the paddle method, the pH values of BCB rapidly increased without the floating lid. The pH change (ΔpH) at 6 h ranged from +1.66 to +1.82 (50 rpm) and +1.96 to +2.02 (100 rpm). The floating lid effectively maintained the pH values in all laboratories, with ΔpH ranging from +0.13 to +0.17 (50 rpm) and +0.21 to +0.25 (100 rpm). The standard deviation of ΔpH was within 0.05 at both 50 and 100 rpm. Similarly, in the FTC method, without the floating lid, ΔpH ranged from +1.71 to +1.77 (reservoir), +1.59 to +1.72 (FTC), and +1.73 to +1.76 (sampling tube). With the floating lid, ΔpH ranged from +0.05 to +0.10 (reservoir), +0.05 to +0.09 (FTC), and +0.26 to +0.39 (sampling tube). The standard deviation of ΔpH was within 0.05. In conclusion, the inter-laboratory reproducibility of pH maintenance of BCB using the floating lid method was confirmed for both the paddle and FTC methods.

Authors

Institutions

Publication Details

Journal
Chemical and Pharmaceutical Bulletin
Published
2025-08-08
DOI
https://doi.org/10.1248/cpb.c25-00283
Citations
2
Primary Topic
Drug Solubulity and Delivery Systems
Type
article
Field-Weighted Citation Impact
1.58
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Bicarbonate Buffer Dissolution Test Using the Floating Lid Method: Inter-Laboratory Reproducibility of pH Maintenance

Masahiro Fushimi, Masaki Higashino, Kiyohiko Sugano, Shotaro Ikuta et al.
2 citations
Chemical and Pharmaceutical Bulletin
Drug Solubulity and Delivery Systems
1.58
article

Bicarbonate Buffer Dissolution Test Using the Floating Lid Method: Inter-Laboratory Reproducibility of pH Maintenance

Masahiro Fushimi, Masaki Higashino, Kiyohiko Sugano, Shotaro Ikuta, Hiroyuki Yoshida
article en
2 citations

Abstract

Bicarbonate buffer (BCB) has been difficult to use in conventional dissolution tests because its pH rapidly increases as CO<sub>2</sub> escapes from the air-water interface. Recently, the floating lid method was introduced as a convenient method for using BCB in dissolution tests. This study aimed to confirm the inter-laboratory reproducibility of pH maintenance of BCB using the floating lid method for both paddle and flow-through cell (FTC) methods. Three pharmaceutical companies and 1 academic research institute participated in this study. A BCB solution (pH 6.5, 15 mM) was employed as the test solution. In the paddle method, the pH values of BCB rapidly increased without the floating lid. The pH change (ΔpH) at 6 h ranged from +1.66 to +1.82 (50 rpm) and +1.96 to +2.02 (100 rpm). The floating lid effectively maintained the pH values in all laboratories, with ΔpH ranging from +0.13 to +0.17 (50 rpm) and +0.21 to +0.25 (100 rpm). The standard deviation of ΔpH was within 0.05 at both 50 and 100 rpm. Similarly, in the FTC method, without the floating lid, ΔpH ranged from +1.71 to +1.77 (reservoir), +1.59 to +1.72 (FTC), and +1.73 to +1.76 (sampling tube). With the floating lid, ΔpH ranged from +0.05 to +0.10 (reservoir), +0.05 to +0.09 (FTC), and +0.26 to +0.39 (sampling tube). The standard deviation of ΔpH was within 0.05. In conclusion, the inter-laboratory reproducibility of pH maintenance of BCB using the floating lid method was confirmed for both the paddle and FTC methods.

Chemical and Pharmaceutical BulletinVol. 73(8)
Ritsumeikan University (JP), Kao Corporation (Japan) (JP), Towa Pharmaceutical (Japan) (JP), National Institute of Health Sciences (JP), Osaka University of Pharmaceutical Sciences (JP)
Clean water and sanitation
Openalex Percentile: Top 15%
Drug Solubulity and Delivery Systems
1.58
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.