Quantitative imaging of 478-keV prompt gamma rays from Boron neutron capture reactions
The accurate imaging and quantitative measurement of 478-keV prompt gamma rays are critical for advancing boron neutron capture therapy (BNCT), a promising cancer treatment. Although numerical simulations have indicated that such measurements are feasible, their practical application has proven challenging. This study introduces a gamma-ray imaging detector designed specifically for precise BNCT measurements. Using boron-rich phantom samples, we successfully imaged 478-keV gamma rays and established a linear correlation between gamma-ray production and boron concentration. Furthermore, applying this technique in a recognized BNCT treatment facility demonstrated the detector's effectiveness in monitoring boron dose distribution during neutron irradiation, both in pre-treatment diagnostics and throughout the treatment process.
Authors
- Atsushi Takada (ORCID: https://orcid.org/0000-0002-2515-5573)
- Tetsuya Mizumoto (ORCID: https://orcid.org/0000-0002-1971-4746)
- T. Tanimori (ORCID: https://orcid.org/0000-0003-4427-0961)
- Shotaro Komura
- Yoshinori Sakurai
Institutions
- Kyoto University (JP)
- Futaba (Japan) (JP)
Publication Details
- Journal
- Scientific Reports
- Published
- 2025-08-11
- DOI
- https://doi.org/10.1038/s41598-025-14212-8
- Primary Topic
- Nuclear Physics and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Ministry of Economy, Trade and Industry