Radiation dose to the eye lens and the shielding efficacy of a protective eye mask in patients with Hodgkin lymphoma undergoing repeated CT examinations

This study evaluated radiation dose to the eye lens and the shielding efficacy of a radiation-protective eye mask (E-Mask) during repeated chest and abdomen computed tomography (CT) scans. Dose measurements were obtained using an anthropomorphic phantom and an optically stimulated luminescence dosemeter (Model TOPAZ, LANDAUER, USA). A 5-y follow-up scenario for Hodgkin lymphoma was simulated by assuming one scan per quarter, with the single-scan dose multiplied by 20 to represent four scans annually over 5-y. Despite the use of low-dose protocols based on national diagnostic reference levels issued by the Korea Disease Control and Prevention Agency, unshielded cumulative lens doses reached 363.89 mSv (right) and 424.84 mSv (left). Using an E-Mask reduced these doses to 243.96 mSv and 240.13 mSv, corresponding to reductions of 32.96% and 43.48%, respectively. These findings support the use of the E-Mask in radiosensitive populations to reduce lens exposure during long-term CT follow-up.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2025-08-08
DOI
https://doi.org/10.1093/rpd/ncaf092
Citations
1
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
1.36
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Radiation dose to the eye lens and the shielding efficacy of a protective eye mask in patients with Hodgkin lymphoma undergoing repeated CT examinations

Hyo-Gi Kim, D. Y. Rhee, Jinseok Yeo, Hee-June Kim et al.
1 citations
Radiation Protection Dosimetry
Radiation Dose and Imaging
1.36
article

Radiation dose to the eye lens and the shielding efficacy of a protective eye mask in patients with Hodgkin lymphoma undergoing repeated CT examinations

Hyo-Gi Kim, D. Y. Rhee, Jinseok Yeo, Hee-June Kim, Il-Hwan Bae
article en
1 citations

Abstract

This study evaluated radiation dose to the eye lens and the shielding efficacy of a radiation-protective eye mask (E-Mask) during repeated chest and abdomen computed tomography (CT) scans. Dose measurements were obtained using an anthropomorphic phantom and an optically stimulated luminescence dosemeter (Model TOPAZ, LANDAUER, USA). A 5-y follow-up scenario for Hodgkin lymphoma was simulated by assuming one scan per quarter, with the single-scan dose multiplied by 20 to represent four scans annually over 5-y. Despite the use of low-dose protocols based on national diagnostic reference levels issued by the Korea Disease Control and Prevention Agency, unshielded cumulative lens doses reached 363.89 mSv (right) and 424.84 mSv (left). Using an E-Mask reduced these doses to 243.96 mSv and 240.13 mSv, corresponding to reductions of 32.96% and 43.48%, respectively. These findings support the use of the E-Mask in radiosensitive populations to reduce lens exposure during long-term CT follow-up.

Radiation Protection DosimetryVol. 201(16)
Kyungpook National University Hospital (KR), Daegu Catholic University (KR)
Good health and well-being
Openalex Percentile: Top 17%
Radiation Dose and Imaging
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Radiation dose to the eye lens and the shielding efficacy of a protective eye mask in patients with Hodgkin lymphoma undergoing repeated CT examinations — Hyo-Gi Kim, D. Y. Rhee, et al. · Radiation Protection Dosimetry (2025) | TGRS Research Map | TGRS