Background Non-neuro arterial embolization is widely performed for various clinical indications, with support systems implemented to ensure procedural safety. However, when the target vessels are highly tortuous or have significant bends at the peripheral level, the microcatheter and coil must traverse lengthy, sharply curved paths, raising concerns about coil maldelivery. Purpose To examine whether sharp vessel angulation (≥90-180°) and microcatheter-coil compatibility independently increase coil maldelivery in non-neuroarterial embolization. Materials and methods This single-center, IRB-approved analysis from February 2023 to December 2024 included 451 arterial branch-detachable coil combinations (BCCs) in 119 patients (mean age 64 years, range 32-87). Angulations of 90-180° at the proximal catheter segment and distal tip were evaluated on digital subtraction angiography. Technical failure was defined as the inability to deploy the coil as intended, including unraveling or coil shape distortion. A 1:3 propensity score matching (28 failed vs 61 successful BCCs), balanced coil features, and target-vessel factors. Mismatch was recorded if the coil's primary diameter exceeded or fell below microcatheter thresholds. A generalized linear mixed model accounted for within-patient clustering. Results Coil failure occurred in 28 of 451 BCCs (6.2 %). A 90-180° inversion at the catheter tip (odds ratio (OR), 22; p = 0.008) and mismatch (OR, 4.9; p = 0.03) independently predicted failure. A proximal 90-180° inversion also contributed (OR, 4.5; p = 0.03). Of 28 failures, 21/28 (75%) were mismatched: 16/21 (76%) resolved via mismatch correction, and 5/21 (24%) required repositioning or an alternate coil gauge/length. Proper-match failures (n=7) were treated with thinner or more flexible coils in 6 (85.7%; p < 0.0001). Conclusions Sharp vessel angulation (≥90-180°) and microcatheter-coil mismatch independently increase the risk of maldelivery in non-neuro arterial embolization.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2025-08-09
DOI
https://doi.org/10.7759/cureus.89699
Primary Topic
Intracranial Aneurysms: Treatment and Complications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Vessel Curvature and Microcatheter-Detachable Coil Compatibility in Arterial Coil Embolization

Junichi Matsumoto, Takahiro Ogi, H. Osanai, Nobuyuki Asato et al.
Cureus
Intracranial Aneurysms: Treatment and Complications
article

Vessel Curvature and Microcatheter-Detachable Coil Compatibility in Arterial Coil Embolization

Junichi Matsumoto, Takahiro Ogi, H. Osanai, Nobuyuki Asato, Fumihito Toshima, Kenichiro Okumura, Kazuto Kozaka, Atsushi Takamatsu, Taichi Kitagawa, Satoshi Kobayashi
article en

Abstract

Background Non-neuro arterial embolization is widely performed for various clinical indications, with support systems implemented to ensure procedural safety. However, when the target vessels are highly tortuous or have significant bends at the peripheral level, the microcatheter and coil must traverse lengthy, sharply curved paths, raising concerns about coil maldelivery. Purpose To examine whether sharp vessel angulation (≥90-180°) and microcatheter-coil compatibility independently increase coil maldelivery in non-neuroarterial embolization. Materials and methods This single-center, IRB-approved analysis from February 2023 to December 2024 included 451 arterial branch-detachable coil combinations (BCCs) in 119 patients (mean age 64 years, range 32-87). Angulations of 90-180° at the proximal catheter segment and distal tip were evaluated on digital subtraction angiography. Technical failure was defined as the inability to deploy the coil as intended, including unraveling or coil shape distortion. A 1:3 propensity score matching (28 failed vs 61 successful BCCs), balanced coil features, and target-vessel factors. Mismatch was recorded if the coil's primary diameter exceeded or fell below microcatheter thresholds. A generalized linear mixed model accounted for within-patient clustering. Results Coil failure occurred in 28 of 451 BCCs (6.2 %). A 90-180° inversion at the catheter tip (odds ratio (OR), 22; p = 0.008) and mismatch (OR, 4.9; p = 0.03) independently predicted failure. A proximal 90-180° inversion also contributed (OR, 4.5; p = 0.03). Of 28 failures, 21/28 (75%) were mismatched: 16/21 (76%) resolved via mismatch correction, and 5/21 (24%) required repositioning or an alternate coil gauge/length. Proper-match failures (n=7) were treated with thinner or more flexible coils in 6 (85.7%; p < 0.0001). Conclusions Sharp vessel angulation (≥90-180°) and microcatheter-coil mismatch independently increase the risk of maldelivery in non-neuro arterial embolization.

CureusVol. 17(8)
Kanazawa University Hospital (JP)
Openalex Percentile: Top 70%
Intracranial Aneurysms: Treatment and Complications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.