An Ultrasensitive Four-Response Ratiometric Electrochemical Homogeneous Biosensor Based on a Positive MOF Biomaterial for Direct Diagnosis of Breast Cancer Types in Human Serum

Continued development of high-performance, low-cost, in vitro diagnostic tools is critical for cancer diagnosis and therapeutic prognosis. Herein, we propose a four-response ratiometric electrochemical homogeneous biosensor based on positive MOF biomaterial for detection of the estrogen receptor (ER), progesterone receptor (PR), human epidermal growth factor receptor-2 (HER2), and cell proliferation index (Ki67) protein biomarkers, and for simultaneous diagnosis of 10 types of breast cancer directly in human serum using a single electrode within one system. The principle of this specific four-response ratiometric electrochemical biosensor relied on neutral red (NR), tolonium blue (TB), tetramethylbenzidine (TMB), and malachite green (MG) loaded on the surface of UiO-66-NH<sub>2</sub>. These electrically active organic molecules with positive charges can be driven by negative voltage-induced diffusion to concentrate on the surface of an ITO electrode via electrostatic interactions. However, if the target bioprotein molecule shows a positive charge when the pH value of the PBS buffer solution is below its isoelectric point, it will change the diffusion path of the electroactive organic molecule due to the charge repulsion, thereby affecting the diffusion rate. The biosensor has a short detection time of 60 min; however, the current clinical tissue testing time takes about 1 week. Additionally, the biosensor selectively detects HER2, ER, Ki67, and PR in a range of 0.0134/0.0331/0.0194/0.0250-35 pg/mL with detection limits of 0.0134, 0.0331, 0.0194, and 0.0250 pg/mL, respectively. Notably, compared with traditional gene chip and immunohistochemistry technology, the test object achieves the transformation from tissue to body fluid analysis by electrochemical sensing technology.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2025-08-11
DOI
https://doi.org/10.1021/acs.analchem.4c06951
Citations
2
Primary Topic
Advanced biosensing and bioanalysis techniques
Type
article
Field-Weighted Citation Impact
1.61

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

An Ultrasensitive Four-Response Ratiometric Electrochemical Homogeneous Biosensor Based on a Positive MOF Biomaterial for Direct Diagnosis of Breast Cancer Types in Human Serum

Yi Ma, Danqun Huo, Changjun Hou, Jingzhou Hou et al.
2 citations
Analytical Chemistry
Advanced biosensing and bioanalysis techniques
1.61
article

An Ultrasensitive Four-Response Ratiometric Electrochemical Homogeneous Biosensor Based on a Positive MOF Biomaterial for Direct Diagnosis of Breast Cancer Types in Human Serum

Yi Ma, Danqun Huo, Changjun Hou, Jingzhou Hou, Ning Li, Mei Yang, Peng Lu, Ya Zhang
article en
2 citations

Abstract

Continued development of high-performance, low-cost, in vitro diagnostic tools is critical for cancer diagnosis and therapeutic prognosis. Herein, we propose a four-response ratiometric electrochemical homogeneous biosensor based on positive MOF biomaterial for detection of the estrogen receptor (ER), progesterone receptor (PR), human epidermal growth factor receptor-2 (HER2), and cell proliferation index (Ki67) protein biomarkers, and for simultaneous diagnosis of 10 types of breast cancer directly in human serum using a single electrode within one system. The principle of this specific four-response ratiometric electrochemical biosensor relied on neutral red (NR), tolonium blue (TB), tetramethylbenzidine (TMB), and malachite green (MG) loaded on the surface of UiO-66-NH<sub>2</sub>. These electrically active organic molecules with positive charges can be driven by negative voltage-induced diffusion to concentrate on the surface of an ITO electrode via electrostatic interactions. However, if the target bioprotein molecule shows a positive charge when the pH value of the PBS buffer solution is below its isoelectric point, it will change the diffusion path of the electroactive organic molecule due to the charge repulsion, thereby affecting the diffusion rate. The biosensor has a short detection time of 60 min; however, the current clinical tissue testing time takes about 1 week. Additionally, the biosensor selectively detects HER2, ER, Ki67, and PR in a range of 0.0134/0.0331/0.0194/0.0250-35 pg/mL with detection limits of 0.0134, 0.0331, 0.0194, and 0.0250 pg/mL, respectively. Notably, compared with traditional gene chip and immunohistochemistry technology, the test object achieves the transformation from tissue to body fluid analysis by electrochemical sensing technology.

Analytical ChemistryVol. 97(32)
Chongqing University (CN), Sichuan University (CN), National Postdoctoral Association (US), Chongqing University of Technology (CN), Sichuan University of Science and Engineering (CN)
National Natural Science Foundation of China, Graduate School, Chongqing University, Fundamental Research Funds for the Central Universities
Good health and well-being
Openalex Percentile: Top 15%
Advanced biosensing and bioanalysis techniques
1.61
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.