Occupational Radiation Protection: Assessment of Bunker Shielding in Radiotherapy Department

Intensity Modulation Radiation Treatment (IMRT) technique increases significantly head leakage and the workload is also affected compared with Conventional treatment. The equivalent dose from medical Linear Accelerator (LINAC) System must be limited to prevent occupational and public exposure to radiation. Therefore, the shielding design must be adequate. The aim of this study was to reassess the shielding of our LINAC room designed for conventional treatment if we planned to treat 30% of the patients with IMRT technique at Aristide Le Dantec hospital. We propose a method to evaluate the equivalent dose by using empirical formulas from the report 151 of the National Council of Radiation Protection (NCRP151). We estimate the transmission factor for all barriers using the wall thicknesses. Equivalent doses for all surrounded areas were calculated and compared with those found during measurement. The doses observed here are below annual dose limit for occupational and public even when IMRT technique will be used for treatment of 30% of the patients in a LINAC room entirely shielded for conventional treatment. The results show also that the shielding was overestimated by at least a factor of 2. Adequacy of secondary shielding will depend on the IMRT patient workload. For conventional facilities that are being assessed for IMRT therapy, existing primary barriers will typically prove adequate. Accurate and conservative shielding design is essential for radiation protection professionals in clinical practice.

Authors

Publication Details

Journal
World Journal of Nuclear Science and Technology
Published
2025-01-01
DOI
https://doi.org/10.4236/wjnst.2025.153008
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Occupational Radiation Protection: Assessment of Bunker Shielding in Radiotherapy Department

Mamadou Moustapha Dieng, Ndèye Arame Boye Faye, Magatte Diagne, Adji Diop
World Journal of Nuclear Science and Technology
Advanced Radiotherapy Techniques
article

Occupational Radiation Protection: Assessment of Bunker Shielding in Radiotherapy Department

Mamadou Moustapha Dieng, Ndèye Arame Boye Faye, Magatte Diagne, Adji Diop
article en

Abstract

Intensity Modulation Radiation Treatment (IMRT) technique increases significantly head leakage and the workload is also affected compared with Conventional treatment. The equivalent dose from medical Linear Accelerator (LINAC) System must be limited to prevent occupational and public exposure to radiation. Therefore, the shielding design must be adequate. The aim of this study was to reassess the shielding of our LINAC room designed for conventional treatment if we planned to treat 30% of the patients with IMRT technique at Aristide Le Dantec hospital. We propose a method to evaluate the equivalent dose by using empirical formulas from the report 151 of the National Council of Radiation Protection (NCRP151). We estimate the transmission factor for all barriers using the wall thicknesses. Equivalent doses for all surrounded areas were calculated and compared with those found during measurement. The doses observed here are below annual dose limit for occupational and public even when IMRT technique will be used for treatment of 30% of the patients in a LINAC room entirely shielded for conventional treatment. The results show also that the shielding was overestimated by at least a factor of 2. Adequacy of secondary shielding will depend on the IMRT patient workload. For conventional facilities that are being assessed for IMRT therapy, existing primary barriers will typically prove adequate. Accurate and conservative shielding design is essential for radiation protection professionals in clinical practice.

World Journal of Nuclear Science and TechnologyVol. 15(03)
Openalex Percentile: Top 73%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.