Optimised circuit design for precise bite force measurement using flexiforce sensors

The precise measurement of bite force is vital in dental diagnostics, particularly for evaluating tooth restorations, prosthetic interventions, and orthodontic treatments. This study presents the calibration and evaluation of the Flexiforce A301 sensor using optimised low-drive voltage circuits to extend its measurement range. Three circuit configurations, a voltage divider, a feedback resistor, and a feedback resistor with a capacitor were designed, simulated using LTspice, and experimentally validated. Results indicate that the configuration incorporating a feedback resistor provides superior linearity and stability, accurately measuring forces up to 100 kg, outperforming other configurations. This advancement enhances the reliability and range of bite force measurements, offering a robust foundation for high-force dental applications.

Authors

Institutions

Publication Details

Journal
Journal of Medical Engineering & Technology
Published
2025-08-09
DOI
https://doi.org/10.1080/03091902.2025.2542270
Citations
1
Primary Topic
Temporomandibular Joint Disorders
Type
article
Field-Weighted Citation Impact
2.14
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimised circuit design for precise bite force measurement using flexiforce sensors

K. B. Bommegowda, Durga Prasad, Preetha J. Shetty, P S Murali et al.
1 citations
Journal of Medical Engineering & Technology
Temporomandibular Joint Disorders
2.14
article

Optimised circuit design for precise bite force measurement using flexiforce sensors

K. B. Bommegowda, Durga Prasad, Preetha J. Shetty, P S Murali, G. R. Pradyumna, M. S. Ravi
article en
1 citations

Abstract

The precise measurement of bite force is vital in dental diagnostics, particularly for evaluating tooth restorations, prosthetic interventions, and orthodontic treatments. This study presents the calibration and evaluation of the Flexiforce A301 sensor using optimised low-drive voltage circuits to extend its measurement range. Three circuit configurations, a voltage divider, a feedback resistor, and a feedback resistor with a capacitor were designed, simulated using LTspice, and experimentally validated. Results indicate that the configuration incorporating a feedback resistor provides superior linearity and stability, accurately measuring forces up to 100 kg, outperforming other configurations. This advancement enhances the reliability and range of bite force measurements, offering a robust foundation for high-force dental applications.

Journal of Medical Engineering & TechnologyVol. 49(8)
Nitte University (IN)
Openalex Percentile: Top 12%
Temporomandibular Joint Disorders
2.14
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Optimised circuit design for precise bite force measurement using flexiforce sensors — K. B. Bommegowda, Durga Prasad, et al. · Journal of Medical Engineering & Technology (2025) | TGRS Research Map | TGRS