From Bench Testing to Virtual Implantation: A Comparative Study Between Poly‐ <scp>l</scp> ‐Lactic Acid and Nickel‐Titanium Braided Stents

Bioresorbable braided stents represent a promising solution for the treatment of peripheral artery disease, providing temporary mechanical support before gradually degrading into biocompatible byproducts. Previous studies have highlighted their lower mechanical performance compared to permanent metallic stents. However, their implantation in lower limb arteries remains unexplored, leaving uncertainty on whether their mechanical performance is sufficient for effective treatment. The aim of the present study was to evaluate the performance of a poly-l-lactic acid (PLLA) braided stent for the treatment of lower limb arteries through in silico analysis and compare it with that of a nickel-titanium (NiTi) device. A finite element (FE) model of the PLLA stent was implemented and validated against experimental bench test data. Subsequently, the mechanical characteristics of the PLLA device were compared to those of a NiTi stent, with identical geometrical features, through FE simulations of two bench tests (i.e., parallel plate compression and crimping tests). Finally, a virtual implantation procedure of both devices in a patient-specific lower limb artery was conducted by FE analysis, accounting for three different arterial wall conditions, to compare the stents' treatment performance. The FE analysis of the bench tests confirmed that the PLLA stent generated much lower force magnitudes than the NiTi device. Moreover, the virtual implantation procedure indicated the limited short-term performance of the PLLA stent for the treatment of peripheral artery disease in terms of risk for permanent deformations, low lumen gain, high values of incomplete stent apposition and a nonuniform distribution of contact pressure on the arterial wall.

Authors

Institutions

Publication Details

Journal
International Journal for Numerical Methods in Biomedical Engineering
Published
2025-08-01
DOI
https://doi.org/10.1002/cnm.70078
Citations
4
Primary Topic
Peripheral Artery Disease Management
Type
article
Field-Weighted Citation Impact
8.03

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

From Bench Testing to Virtual Implantation: A Comparative Study Between Poly‐ l ‐Lactic Acid and Nickel‐Titanium Braided Stents

Alissa Zaccaria, Thomas Gries, Anna Corti, Dario Carbonaro et al.
4 citations
International Journal for Numerical Methods in Biomedical Engineering
Peripheral Artery Disease Management
8.03
article

From Bench Testing to Virtual Implantation: A Comparative Study Between Poly‐ l ‐Lactic Acid and Nickel‐Titanium Braided Stents

Alissa Zaccaria, Thomas Gries, Anna Corti, Dario Carbonaro, Ted J. Vaughan, Agnese Lucchetti, Claudio Chiastra, Levi G. Juhl
article en
4 citations

Abstract

Bioresorbable braided stents represent a promising solution for the treatment of peripheral artery disease, providing temporary mechanical support before gradually degrading into biocompatible byproducts. Previous studies have highlighted their lower mechanical performance compared to permanent metallic stents. However, their implantation in lower limb arteries remains unexplored, leaving uncertainty on whether their mechanical performance is sufficient for effective treatment. The aim of the present study was to evaluate the performance of a poly-l-lactic acid (PLLA) braided stent for the treatment of lower limb arteries through in silico analysis and compare it with that of a nickel-titanium (NiTi) device. A finite element (FE) model of the PLLA stent was implemented and validated against experimental bench test data. Subsequently, the mechanical characteristics of the PLLA device were compared to those of a NiTi stent, with identical geometrical features, through FE simulations of two bench tests (i.e., parallel plate compression and crimping tests). Finally, a virtual implantation procedure of both devices in a patient-specific lower limb artery was conducted by FE analysis, accounting for three different arterial wall conditions, to compare the stents' treatment performance. The FE analysis of the bench tests confirmed that the PLLA stent generated much lower force magnitudes than the NiTi device. Moreover, the virtual implantation procedure indicated the limited short-term performance of the PLLA stent for the treatment of peripheral artery disease in terms of risk for permanent deformations, low lumen gain, high values of incomplete stent apposition and a nonuniform distribution of contact pressure on the arterial wall.

International Journal for Numerical Methods in Biomedical EngineeringVol. 41(8)
University of Bergamo (IT), Politecnico di Torino (IT), Ollscoil na Gaillimhe – University of Galway (IE), FEG Textiltechnik (Germany) (DE), Bioengineering Technology and Systems (Italy) (IT), RWTH Aachen University (DE)
Boston Scientific Corporation, European Commission, RWTH Aachen University, Horizon 2020, HORIZON EUROPE Marie Sklodowska-Curie Actions
Good health and well-being
Openalex Percentile: Top 2%
Peripheral Artery Disease Management
8.03
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.