Verification of Pathologic ECV Estimated with Clinical T1 Mapping Methods Using an Isotope-based Reference

Purpose To verify the accuracy of clinically used cardiac MRI T1 mapping sequences for measuring extracellular volume fraction (ECV) throughout the spectrum of pathologic values using an independent reference standard. Materials and Methods Acute myocardial ischemia was induced in six pigs using endovascular balloon occlusion in the left anterior descending artery between October 2018 and November 2018. After 7 days of reperfusion, ECV was measured in vivo using four sequences: three modified Look-Locker inversion recovery (MOLLI) and one saturation recovery single-shot acquisition (SASHA) sequences. ECV was also measured using ex vivo SPECT following injection of technetium 99m diethylenetriamine pentaacetic acid. ECV was calculated for corresponding regions of interest of necrosis on cardiac MR and SPECT images and in myocardium remote from injury. Mixed model analysis was used to test for correlation, while the paired <i>t</i> test was used to test for differences in T1 and ECV. Results ECV measurements with all T1 mapping sequences showed high correlation to the reference standard (<i>r<sup>2</sup></i> range, 0.71-0.99). Measurements with MOLLI 5(3)3 and MOLLI 5s(3s)3s overestimated ECV in the infarct region compared with the reference standard (bias ± 2 SDs: 8.2% ± 8.5 and 4.9% ± 8.7; <i>P</i> = .005 and <i>P</i> = .04, respectively). All sequences showed overestimation of ECV in the remote region (range of bias, 3%-14% points, all MOLLI <i>P</i> values ≤ .001, SASHA <i>P</i> = .04). Conclusion This study verified the accuracy of clinically used T1 mapping sequences in measuring ECV in absolute values over a spectrum of pathologic values, using an independent radioisotope-based three-dimensionally acquired reference standard. <b>Keywords:</b> Animal Studies, MR Imaging, Cardiac, Edema, Imaging Sequences, Ischemia/Infarction <i>Supplemental material is available for this article.</i> © RSNA, 2025.

Authors

Institutions

Publication Details

Journal
Radiology Cardiothoracic Imaging
Published
2025-08-01
DOI
https://doi.org/10.1148/ryct.240441
Primary Topic
Viral Infections and Immunology Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Verification of Pathologic ECV Estimated with Clinical T1 Mapping Methods Using an Isotope-based Reference

Sascha Kopic, Håkan Arheden, Kelvin Chow, Robert Jablonowski et al.
Radiology Cardiothoracic Imaging
Viral Infections and Immunology Research
article

Verification of Pathologic ECV Estimated with Clinical T1 Mapping Methods Using an Isotope-based Reference

Sascha Kopic, Håkan Arheden, Kelvin Chow, Robert Jablonowski, David Nordlund, Anthony H. Aletras, Christos G. Xanthis, Mikael Kanski, Jonathan Berg, Sebastian Bidhult
article en

Abstract

Purpose To verify the accuracy of clinically used cardiac MRI T1 mapping sequences for measuring extracellular volume fraction (ECV) throughout the spectrum of pathologic values using an independent reference standard. Materials and Methods Acute myocardial ischemia was induced in six pigs using endovascular balloon occlusion in the left anterior descending artery between October 2018 and November 2018. After 7 days of reperfusion, ECV was measured in vivo using four sequences: three modified Look-Locker inversion recovery (MOLLI) and one saturation recovery single-shot acquisition (SASHA) sequences. ECV was also measured using ex vivo SPECT following injection of technetium 99m diethylenetriamine pentaacetic acid. ECV was calculated for corresponding regions of interest of necrosis on cardiac MR and SPECT images and in myocardium remote from injury. Mixed model analysis was used to test for correlation, while the paired <i>t</i> test was used to test for differences in T1 and ECV. Results ECV measurements with all T1 mapping sequences showed high correlation to the reference standard (<i>r<sup>2</sup></i> range, 0.71-0.99). Measurements with MOLLI 5(3)3 and MOLLI 5s(3s)3s overestimated ECV in the infarct region compared with the reference standard (bias ± 2 SDs: 8.2% ± 8.5 and 4.9% ± 8.7; <i>P</i> = .005 and <i>P</i> = .04, respectively). All sequences showed overestimation of ECV in the remote region (range of bias, 3%-14% points, all MOLLI <i>P</i> values ≤ .001, SASHA <i>P</i> = .04). Conclusion This study verified the accuracy of clinically used T1 mapping sequences in measuring ECV in absolute values over a spectrum of pathologic values, using an independent radioisotope-based three-dimensionally acquired reference standard. <b>Keywords:</b> Animal Studies, MR Imaging, Cardiac, Edema, Imaging Sequences, Ischemia/Infarction <i>Supplemental material is available for this article.</i> © RSNA, 2025.

Radiology Cardiothoracic ImagingVol. 7(4)
Aristotle University of Thessaloniki (GR), Skåne University Hospital (SE), Siemens Healthcare (United States) (US)
Openalex Percentile: Top 68%
Viral Infections and Immunology Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.