Abstract Introduction: AI-based autocontouring products claim to be able to segment organs with accuracy comparable to humans. We compare the geometric and dosimetric performance of three AI-based autocontouring packages (Autocontour 2.5.6, (“RF”); Annotate 2.3.1, (“TP”) and RT-Mind_AI 1.0, (“MM”)) in the head and neck region. Methods: We generated 14 organ at risk (OAR) autocontours on 13 computed tomography (CT) image sets. They were compared with clinical (human-generated) contours. The geometric differences were quantified by calculating Dice coefficients and Hausdorff distances. The autocontours were compared visually with the clinical controus by an expert physician. The autocontour sets were also ranked for accuracy by two physicians. The dosimetric effects were evaluated by recalculating treatment plans on the autocontoured CT sets. Results: RF and TP slightly outperformed MM in geometric metrics (the percentage of OARs having mean Dice coefficients > 0.7 was RF 57.1 %, TP 64.3 % and MM 50.0%). The physician judged RF and TP contours to be more anatomically accurate, on average, than the manual contours (manual contour mean accuracy score 2.49, RF 2.28, MM 3.24, TP 1.93). The mean scores given to the autocontours by the two physicians were better for RF and TP, compared to MM (RF 1.86, MM 2.36, TP 1.77). The dosimetric differences were similar for all three programs and were not strongly correlated with the geometric differences. Conclusions: The performance of the three autocontouring packages in the head and neck region is similar, with TP and RF slightly outperforming MM. The correlation between geometric and dosimetric metrics is not strong, and dosimetric evaluation is therefore recommended before clinical use of autocontouring software.

Authors

Institutions

Publication Details

Journal
Journal of Radiotherapy in Practice
Published
2025-01-01
DOI
https://doi.org/10.1017/s1460396925100204
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A geometric and dosimetric comparison of three AI-based autocontouring packages in the head and neck region

Jussi Sillanpaa, Margaret Reynolds, Amit J. Sood
Journal of Radiotherapy in Practice
Advanced Radiotherapy Techniques
article

A geometric and dosimetric comparison of three AI-based autocontouring packages in the head and neck region

Jussi Sillanpaa, Margaret Reynolds, Amit J. Sood
article en

Abstract

Abstract Introduction: AI-based autocontouring products claim to be able to segment organs with accuracy comparable to humans. We compare the geometric and dosimetric performance of three AI-based autocontouring packages (Autocontour 2.5.6, (“RF”); Annotate 2.3.1, (“TP”) and RT-Mind_AI 1.0, (“MM”)) in the head and neck region. Methods: We generated 14 organ at risk (OAR) autocontours on 13 computed tomography (CT) image sets. They were compared with clinical (human-generated) contours. The geometric differences were quantified by calculating Dice coefficients and Hausdorff distances. The autocontours were compared visually with the clinical controus by an expert physician. The autocontour sets were also ranked for accuracy by two physicians. The dosimetric effects were evaluated by recalculating treatment plans on the autocontoured CT sets. Results: RF and TP slightly outperformed MM in geometric metrics (the percentage of OARs having mean Dice coefficients > 0.7 was RF 57.1 %, TP 64.3 % and MM 50.0%). The physician judged RF and TP contours to be more anatomically accurate, on average, than the manual contours (manual contour mean accuracy score 2.49, RF 2.28, MM 3.24, TP 1.93). The mean scores given to the autocontours by the two physicians were better for RF and TP, compared to MM (RF 1.86, MM 2.36, TP 1.77). The dosimetric differences were similar for all three programs and were not strongly correlated with the geometric differences. Conclusions: The performance of the three autocontouring packages in the head and neck region is similar, with TP and RF slightly outperforming MM. The correlation between geometric and dosimetric metrics is not strong, and dosimetric evaluation is therefore recommended before clinical use of autocontouring software.

Journal of Radiotherapy in PracticeVol. 24
University of Minnesota System (US)
Openalex Percentile: Top 73%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A geometric and dosimetric comparison of three AI-based autocontouring packages in the head and neck region — Jussi Sillanpaa, Margaret Reynolds, et al. · Journal of Radiotherapy in Practice (2025) | TGRS Research Map | TGRS