A comparison of electrophysiological measures for characterizing the cochlear-implant electrode-neuron interface
The health of the auditory nerve is a key factor for hearing outcomes in cochlear-implant patients. Many electrophysiological markers of auditory-neural health have been proposed, with varying predictive power and independence from non-neural factors. The "Failure Index" (FI), the ratio between stimulation current level and response magnitude of the electrically-evoked compound action potential (eCAP) was compared to the Panoramic eCAP method. Both methods predicted localized areas of reduced neural responsiveness in a group of human cochlear-implant users, but the FI showed a greater dependence on a non-neural factor, namely, the distances between the cochlear-implant electrodes and the auditory nerves.
Authors
- Fergio Sismono (ORCID: https://orcid.org/0000-0002-4611-8632)
- Dorothée Arzounian (ORCID: https://orcid.org/0000-0003-1520-0227)
- Charlotte Garcia (ORCID: https://orcid.org/0000-0003-0370-3927)
- François Guérit (ORCID: https://orcid.org/0000-0002-9404-3575)
- Robert P. Carlyon (ORCID: https://orcid.org/0000-0002-6166-501X)
- Tobias Goehring (ORCID: https://orcid.org/0000-0002-9038-3310)
Institutions
- MRC Cognition and Brain Sciences Unit (GB)
- MRC Epidemiology Unit (GB)
Publication Details
- Journal
- JASA Express Letters
- Published
- 2025-08-01
- DOI
- https://doi.org/10.1121/10.0038746
- Citations
- 4
- Primary Topic
- Hearing Loss and Rehabilitation
- Type
- article
- Field-Weighted Citation Impact
- 5.72
Funders
- Wellcome Trust
- Royal National Institute for Deaf People
- Medical Research Council