Sparge Sampling of Molten Salts for Online Monitoring via Laser-Induced Breakdown Spectroscopy

A method was developed to sample molten salts by sparging to generate and transport aerosols to an isolated instrument for compositional analysis by laser-induced breakdown spectroscopy (LIBS). Real-time monitoring of molten salt composition is critical to developing molten salt nuclear reactors, which offer enhanced safety and efficiency. In this article, the sparge sampling method is described and compared with sampling using a Collison nebulizer. The size distribution and transport of aerosols produced from molten eutectic NaNO<sub>3</sub>-KNO<sub>3</sub> salt were compared for multiple gas flow rates (75-1200 mL min<sup>-1</sup>) and transport distances (0.68-2.61 m). Both methods produced aerosols ranging from 0.5 to 5.0 μm determined using a cascade impactor. Aerosols were effectively transported without pre- or trace-heating of gas lines, but transport efficiency was reduced by the formation of agglomerates. Sparge sampling was found to use less sample and less gas than a Collison nebulizer while producing a more concentrated aerosol stream (up to 5 μg L<sup>-1</sup>). The effects of laser energy and delay time on the signal quality of LIBS measurements of these aerosols were also studied. High energy and short delay times were found to enhance signal and repeatability, whereas signal-to-background and signal-to-noise ratios were highest at low energy and longer delay times. The capabilities of this system for online monitoring of molten salts were demonstrated with calibrations for Sr and Li with relative standard deviations of 2.6% and 1.5% and limits of detection of 380 and 180 μg g<sup>-1</sup>, respectively.

Authors

Institutions

Publication Details

Journal
ACS Omega
Published
2025-08-05
DOI
https://doi.org/10.1021/acsomega.5c04988
Citations
4
Primary Topic
Laser-induced spectroscopy and plasma
Type
article
Field-Weighted Citation Impact
3.45

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Sparge Sampling of Molten Salts for Online Monitoring via Laser-Induced Breakdown Spectroscopy

Joanna McFarlane, Benjamin T. Manard, Zechariah B. Kitzhaber, Hunter B. Andrews et al.
4 citations
ACS Omega
Laser-induced spectroscopy and plasma
3.45
article

Sparge Sampling of Molten Salts for Online Monitoring via Laser-Induced Breakdown Spectroscopy

Joanna McFarlane, Benjamin T. Manard, Zechariah B. Kitzhaber, Hunter B. Andrews, Daniel Orea
article en
4 citations

Abstract

A method was developed to sample molten salts by sparging to generate and transport aerosols to an isolated instrument for compositional analysis by laser-induced breakdown spectroscopy (LIBS). Real-time monitoring of molten salt composition is critical to developing molten salt nuclear reactors, which offer enhanced safety and efficiency. In this article, the sparge sampling method is described and compared with sampling using a Collison nebulizer. The size distribution and transport of aerosols produced from molten eutectic NaNO<sub>3</sub>-KNO<sub>3</sub> salt were compared for multiple gas flow rates (75-1200 mL min<sup>-1</sup>) and transport distances (0.68-2.61 m). Both methods produced aerosols ranging from 0.5 to 5.0 μm determined using a cascade impactor. Aerosols were effectively transported without pre- or trace-heating of gas lines, but transport efficiency was reduced by the formation of agglomerates. Sparge sampling was found to use less sample and less gas than a Collison nebulizer while producing a more concentrated aerosol stream (up to 5 μg L<sup>-1</sup>). The effects of laser energy and delay time on the signal quality of LIBS measurements of these aerosols were also studied. High energy and short delay times were found to enhance signal and repeatability, whereas signal-to-background and signal-to-noise ratios were highest at low energy and longer delay times. The capabilities of this system for online monitoring of molten salts were demonstrated with calibrations for Sr and Li with relative standard deviations of 2.6% and 1.5% and limits of detection of 380 and 180 μg g<sup>-1</sup>, respectively.

ACS OmegaVol. 10(33)
Oak Ridge National Laboratory (US)
Office of Nuclear Energy
Openalex Percentile: Top 7%
Laser-induced spectroscopy and plasma
3.45
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.