Evaluation of the safety and performance of the HeartTouch™ Electrophysiology Mapping Catheter and NaviBlaze™ Electrophysiology Ablation Catheter in a porcine model

The objective of this study was to evaluate the safety and performance of the HeartTouch™ Electrophysiology Mapping Catheter and NaviBlaze™ Electrophysiology Ablation Catheter in various heart chambers (right atrium, right ventricle, left atrium, and pulmonary veins) in a porcine model. The study utilized two female pigs. The procedures were carried out under anesthesia with standard percutaneous access through the 7F sheath which was inserted into the internal jugular vein, two sheaths of 8F into the right femoral vein and one 5F sheath into the left femoral artery. The study involved multiple runs of mapping and ablation catheter tests to assess both mapping accuracy and lesion effectiveness. Mapping catheter testing demonstrated successful navigation and signal recording across various heart chambers, confirming the efficacy of device in identifying electrophysiological landmarks. Ablation catheter testing involved both non-irrigated and irrigated catheters, resulting in significant impedance reductions at target sites, confirming effective tissue interaction and lesion creation. Necropsy images revealed visible ablation lesions, further corroborating the catheter's effectiveness. No morbidity or mortality was observed in the animals, and clinical signs remained normal throughout the study. The findings suggest that both the HeartTouch™ Mapping Catheter and NaviBlaze™ Ablation Catheter are effective tools for precise mapping and targeted ablation, supporting their potential for clinical use in treating arrhythmias and advancing electrophysiological research

Authors

Institutions

Publication Details

Journal
International Journal of Allied Medical Sciences and Clinical Research
Published
2025-06-06
DOI
https://doi.org/10.61096/ijamscr.v13.iss2.2025.280-296
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of the safety and performance of the HeartTouch™ Electrophysiology Mapping Catheter and NaviBlaze™ Electrophysiology Ablation Catheter in a porcine model

Minocha Dr. Pramodkumar, S. Divya, P. Milan, Pandya Kamna et al.
1 links
International Journal of Allied Medical Sciences and Clinical Research
Cardiac Arrhythmias and Treatments
article

Evaluation of the safety and performance of the HeartTouch™ Electrophysiology Mapping Catheter and NaviBlaze™ Electrophysiology Ablation Catheter in a porcine model

Minocha Dr. Pramodkumar, S. Divya, P. Milan, Pandya Kamna, Dave Arpit, Kothwala Dr. Deveshkumar
article en

Abstract

The objective of this study was to evaluate the safety and performance of the HeartTouch™ Electrophysiology Mapping Catheter and NaviBlaze™ Electrophysiology Ablation Catheter in various heart chambers (right atrium, right ventricle, left atrium, and pulmonary veins) in a porcine model. The study utilized two female pigs. The procedures were carried out under anesthesia with standard percutaneous access through the 7F sheath which was inserted into the internal jugular vein, two sheaths of 8F into the right femoral vein and one 5F sheath into the left femoral artery. The study involved multiple runs of mapping and ablation catheter tests to assess both mapping accuracy and lesion effectiveness. Mapping catheter testing demonstrated successful navigation and signal recording across various heart chambers, confirming the efficacy of device in identifying electrophysiological landmarks. Ablation catheter testing involved both non-irrigated and irrigated catheters, resulting in significant impedance reductions at target sites, confirming effective tissue interaction and lesion creation. Necropsy images revealed visible ablation lesions, further corroborating the catheter's effectiveness. No morbidity or mortality was observed in the animals, and clinical signs remained normal throughout the study. The findings suggest that both the HeartTouch™ Mapping Catheter and NaviBlaze™ Ablation Catheter are effective tools for precise mapping and targeted ablation, supporting their potential for clinical use in treating arrhythmias and advancing electrophysiological research

International Journal of Allied Medical Sciences and Clinical ResearchVol. 13(2)
Meril Life Sciences (India) (IN)
Good health and well-being
Openalex Percentile: Top 69%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.