Stylet-driven lead fracture in left bundle branch area pacing: a case report

Background: Left bundle branch area pacing (LBBAP) represents an advanced physiologic conduction system pacing technique.However, the long-term stability of stylet-driven leads (SDLs) in LBBAP remains uncertain, with limited documented reports on lead fractures.Case presentation: An 81-year-old male with a history of chronic kidney disease, heart failure, and paroxysmal atrial fibrillation (AF) was admitted for medical management.The patient experienced persistent dyspnea and dizziness attributable to bradycardia and required ongoing rhythm control medication for paroxysmal AF.Given the presence of an underlying left bundle branch block with a mildly reduced left ventricular ejection fraction, a dual-chamber pacemaker (Enitra 8 DR-T; Biotronik, DDDR) with LBBAP using SDL (Solia S50) was implanted.As the patient underwent dialysis via the left arm, pacemaker implantation was performed through the right axillary vein.At 30 months post-implantation, elevated lead impedance exceeding 2,000 and complete ventricular lead dysfunction were highly suggestive of lead fracture despite fluoroscopic imaging showing no definitive evidence of fracture.During ventricular lead extraction, severe adhesions and right axillary vein obstruction were encountered, requiring an 11Fr TightRail device for successful removal.The extracted lead displayed a distinct bend near the interelectrode segment, confirming a typical SDL fracture in LBBAP.Conclusions: This case highlights the typical characteristics of lead fracture and precautions associated with SDL-LBBAP and emphasizes that such fractures can occur beyond previously reported timeframes.To prevent SDL-LBBAP fracture, careful attention to lead manipulation during rotational adjustments is essential during the procedure.

Authors

Institutions

Publication Details

Journal
International Journal of Arrhythmia
Published
2025-01-01
DOI
https://doi.org/10.18501/ija.2025.26.e4
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Stylet-driven lead fracture in left bundle branch area pacing: a case report

Kyung‐Yeon Lee, Eue‐Keun Choi, Seil Oh, Hyo‐Jeong Ahn et al.
1 links
International Journal of Arrhythmia
Cardiac pacing and defibrillation studies
article

Stylet-driven lead fracture in left bundle branch area pacing: a case report

Kyung‐Yeon Lee, Eue‐Keun Choi, Seil Oh, Hyo‐Jeong Ahn, So‐Ryoung Lee, Jae-Hyun Kim
article en

Abstract

Background: Left bundle branch area pacing (LBBAP) represents an advanced physiologic conduction system pacing technique.However, the long-term stability of stylet-driven leads (SDLs) in LBBAP remains uncertain, with limited documented reports on lead fractures.Case presentation: An 81-year-old male with a history of chronic kidney disease, heart failure, and paroxysmal atrial fibrillation (AF) was admitted for medical management.The patient experienced persistent dyspnea and dizziness attributable to bradycardia and required ongoing rhythm control medication for paroxysmal AF.Given the presence of an underlying left bundle branch block with a mildly reduced left ventricular ejection fraction, a dual-chamber pacemaker (Enitra 8 DR-T; Biotronik, DDDR) with LBBAP using SDL (Solia S50) was implanted.As the patient underwent dialysis via the left arm, pacemaker implantation was performed through the right axillary vein.At 30 months post-implantation, elevated lead impedance exceeding 2,000 and complete ventricular lead dysfunction were highly suggestive of lead fracture despite fluoroscopic imaging showing no definitive evidence of fracture.During ventricular lead extraction, severe adhesions and right axillary vein obstruction were encountered, requiring an 11Fr TightRail device for successful removal.The extracted lead displayed a distinct bend near the interelectrode segment, confirming a typical SDL fracture in LBBAP.Conclusions: This case highlights the typical characteristics of lead fracture and precautions associated with SDL-LBBAP and emphasizes that such fractures can occur beyond previously reported timeframes.To prevent SDL-LBBAP fracture, careful attention to lead manipulation during rotational adjustments is essential during the procedure.

International Journal of ArrhythmiaVol. 26(1)
Seoul National University (KR), Seoul National University Hospital (KR)
Climate action
Openalex Percentile: Top 69%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.