Evaluating dose uncertainty that is caused by excess CBCT frames on Elekta XVI systems using log file analysis – technical note

Abstract Introduction: This report investigates excess frame count during radiotherapy sessions using Elekta Versa HD systems with X-ray Volume Imaging (XVI) technology at Singleton Hospital. The hospital has 4 clinical linear accelerators (linacs) with XVI, which were analysed to identity variations in the number of excess frames between machines and imaging protocols. Such deviations could affect imaging dose accuracy, patient safety, and system efficiency. Method: XVI log files were gathered from each linac over an 18-month period using data backups. A Python script was created to read and link all the required data in a simple format to generate histograms and tables. Results: The excessive frames resulted in increased radiation doses. Although individual doses were negligible, the highest excess dose for a single patient was 0.7 mSv in 1 fraction, leading to a total dose of 3.4 mSv instead of the expected 2.7 mSv scans which is equivalent to 3 months of background radiation extra. The study revealed that 1.7 % of all imaging sessions were affected (417 imaging fractions). It was identified that the ‘Fast’ Breast imaging protocols were more likely to generate excess frames, likely to be due to the increased gantry speed. Conclusion: Despite the small individual doses, the findings raise concerns about system performance and patient safety for imaging, emphasising the need for further investigation to ensure optimal treatment accuracy and compliance with the Ionising Radiation (Medical Exposure) (Amendment) 2024), Regulation 11 and 12.

Authors

Publication Details

Journal
Journal of Radiotherapy in Practice
Published
2025-01-01
DOI
https://doi.org/10.1017/s1460396925100198
Citations
1
Primary Topic
Nuclear reactor physics and engineering
Type
article
Field-Weighted Citation Impact
4.99
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluating dose uncertainty that is caused by excess CBCT frames on Elekta XVI systems using log file analysis – technical note

R. B. Jenkins, O. Dale Williams, Mark Edwards, Ryan Lewis et al.
1 citations
Journal of Radiotherapy in Practice
Nuclear reactor physics and engineering
4.99
article

Evaluating dose uncertainty that is caused by excess CBCT frames on Elekta XVI systems using log file analysis – technical note

R. B. Jenkins, O. Dale Williams, Mark Edwards, Ryan Lewis, Artjoms Smakovs
article en
1 citations

Abstract

Abstract Introduction: This report investigates excess frame count during radiotherapy sessions using Elekta Versa HD systems with X-ray Volume Imaging (XVI) technology at Singleton Hospital. The hospital has 4 clinical linear accelerators (linacs) with XVI, which were analysed to identity variations in the number of excess frames between machines and imaging protocols. Such deviations could affect imaging dose accuracy, patient safety, and system efficiency. Method: XVI log files were gathered from each linac over an 18-month period using data backups. A Python script was created to read and link all the required data in a simple format to generate histograms and tables. Results: The excessive frames resulted in increased radiation doses. Although individual doses were negligible, the highest excess dose for a single patient was 0.7 mSv in 1 fraction, leading to a total dose of 3.4 mSv instead of the expected 2.7 mSv scans which is equivalent to 3 months of background radiation extra. The study revealed that 1.7 % of all imaging sessions were affected (417 imaging fractions). It was identified that the ‘Fast’ Breast imaging protocols were more likely to generate excess frames, likely to be due to the increased gantry speed. Conclusion: Despite the small individual doses, the findings raise concerns about system performance and patient safety for imaging, emphasising the need for further investigation to ensure optimal treatment accuracy and compliance with the Ionising Radiation (Medical Exposure) (Amendment) 2024), Regulation 11 and 12.

Journal of Radiotherapy in PracticeVol. 24
Openalex Percentile: Top 5%
Nuclear reactor physics and engineering
4.99
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.