Double QRS Transition Due to Anodal Capture During Left Bundle Branch Area Pacing: A Case Report
Anodal capture, characterized by a different QRS morphology compared to cathodal capture, is a well-known issue in cardiac resynchronization therapy (CRT). Left bundle branch area pacing (LBBAP), a novel physiological pacing technique, is also used as a bailout strategy following failed conventional CRT implantation. In LBBAP, QRS transition, defined by a change in paced QRS morphology, serves as a key marker of successful lead placement. This case report is the first to document both high-output anodal capture and LBBAP-induced QRS transition in a single individual receiving LBBAP with an implantable cardioverter-defibrillator (ICD) as a bailout strategy for failed cardiac resynchronization therapy with defibrillator (CRT-D) implantation. Their coexistence underscores unique device optimization challenges in this emerging approach.
Authors
- Luca Canovi (ORCID: https://orcid.org/0009-0005-5006-156X)
- Matteo Bertini (ORCID: https://orcid.org/0000-0002-5285-7140)
- Francesco Vitali (ORCID: https://orcid.org/0000-0002-0667-7616)
- Angelo Melpignano
- L Vocale
- Jacopo Bonini
Institutions
- Arcispedale Sant'Anna (IT)
Publication Details
- Journal
- Journal of Cardiovascular Development and Disease
- Published
- 2025-08-03
- DOI
- https://doi.org/10.3390/jcdd12080299
- Primary Topic
- Cardiac pacing and defibrillation studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00