Comparative Analysis of Transcutaneous Bilirubin Measurement Techniques and Their Agreement With Serum Bilirubin Levels in Neonatal Jaundice: A Prospective Study

Aim: This prospective study aimed to assess the accuracy and agreement of two transcutaneous bilirubin (TcB) devices—BiliCare and MBJ20—compared with serum total bilirubin (STB) in neonates with jaundice. Methods: Eighty‐eight infants born at ≥ 34 weeks of gestation were included. TcB was measured from the auricle using BiliCare and from the sternum using MBJ20, concurrently with STB testing. Results: The median gestational age was 38 weeks and mean birth weight was 3108 g; 39% were delivered by cesarean section and 49% were male. Mean bilirubin levels were 14.6 mg/dL (MBJ20), 12.9 mg/dL (BiliCare), and 14.4 mg/dL (STB). Both devices showed strong correlation with STB (MBJ20: r = 0.756, BiliCare: r = 0.772; p < 0.001). Mean differences from STB were −0.12 mg/dL for MBJ20 and 1.54 mg/dL for BiliCare. BiliCare showed higher predictive accuracy (AUC = 0.814) than MBJ20 (AUC = 0.751). Optimal TcB cutoff values were 14.2 mg/dL (BiliCare) and 15.6 mg/dL (MBJ20). Regression analysis showed no significant bias for either device. Based on STB measurements, 15% of the infants required phototherapy (PT). Conclusion: Both devices provide reliable noninvasive screening though BiliCare has a better accuracy in predicting the need for PT.

Authors

Institutions

Publication Details

Journal
International Journal of Clinical Practice
Published
2025-01-01
DOI
https://doi.org/10.1155/ijcp/1688130
Primary Topic
Neonatal Health and Biochemistry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative Analysis of Transcutaneous Bilirubin Measurement Techniques and Their Agreement With Serum Bilirubin Levels in Neonatal Jaundice: A Prospective Study

Merve Küçükoğlu Keser, Fatma Nur Sarı, Esin Okman, Esra Beşer et al.
International Journal of Clinical Practice
Neonatal Health and Biochemistry
article

Comparative Analysis of Transcutaneous Bilirubin Measurement Techniques and Their Agreement With Serum Bilirubin Levels in Neonatal Jaundice: A Prospective Study

Merve Küçükoğlu Keser, Fatma Nur Sarı, Esin Okman, Esra Beşer, Evrim Alyamaç Dizdar
article en

Abstract

Aim: This prospective study aimed to assess the accuracy and agreement of two transcutaneous bilirubin (TcB) devices—BiliCare and MBJ20—compared with serum total bilirubin (STB) in neonates with jaundice. Methods: Eighty‐eight infants born at ≥ 34 weeks of gestation were included. TcB was measured from the auricle using BiliCare and from the sternum using MBJ20, concurrently with STB testing. Results: The median gestational age was 38 weeks and mean birth weight was 3108 g; 39% were delivered by cesarean section and 49% were male. Mean bilirubin levels were 14.6 mg/dL (MBJ20), 12.9 mg/dL (BiliCare), and 14.4 mg/dL (STB). Both devices showed strong correlation with STB (MBJ20: r = 0.756, BiliCare: r = 0.772; p < 0.001). Mean differences from STB were −0.12 mg/dL for MBJ20 and 1.54 mg/dL for BiliCare. BiliCare showed higher predictive accuracy (AUC = 0.814) than MBJ20 (AUC = 0.751). Optimal TcB cutoff values were 14.2 mg/dL (BiliCare) and 15.6 mg/dL (MBJ20). Regression analysis showed no significant bias for either device. Based on STB measurements, 15% of the infants required phototherapy (PT). Conclusion: Both devices provide reliable noninvasive screening though BiliCare has a better accuracy in predicting the need for PT.

International Journal of Clinical PracticeVol. 2025(1)
Bilkent University (TR), Memorial Ankara Hospital (TR), Başkent University Hospital (TR), Sağlık Bilimleri Üniversitesi (TR)
Openalex Percentile: Top 73%
Neonatal Health and Biochemistry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.