Objective: To assess the diagnostic accuracy of transcutaneous bilirubinometry over serum bilirubin level in newborns with hyperbilirubinemia. Methodology: This cross-sectional study was conducted at Department of Neonatology, Children Hospital, PIMS Islamabad for six months from 01-09-20 to 28-02-21, included 150 jaundiced neonates of any gender over six months. Exclusion criteria included pre-treated hyperbilirubinemia, very premature or low birth weight infants, congenital abnormalities, or infants of jaundiced mothers. Post ethical clearance and parental consent, neonates were evaluated using the MBJ-20 transcutaneous bilirubinometer, with results compared to serum bilirubin levels. The study identified true and false diagnostic outcomes based on a ?12 mg/dL bilirubin threshold, signifying phototherapy initiation. Data analysis was performed with SPSS 26.0, calculating sensitivity, specificity, and accuracy, and determining diagnostic congruence through Pearson’s correlation and ROC AUC, with a significance threshold at p<0.05. Results: The mean serum bilirubin level and transcutaneous bilirubin levels were 14.28 ± 2.21 mg/dL and 14.52 ± 2.72 mg/dL respectively, with a mean difference of 0.24± 1.38 mg/dL (t(149) = 2.141, p = .034). This difference demonstrating a statistically significant overestimation compared to serum bilirubin levels. The correlation between transcutaneous and serum bilirubin levels was strong (r = 0.864, p < 0.0001). For the 12 mg/dL phototherapy threshold, the transcutaneous method showed high sensitivity (95.59%) and specificity (78.57%), with an overall test accuracy of 94%. The Area Under the Curve (AUC) for the receiver operating characteristic was 0.962, indicating high diagnostic precision. Conclusion: Transcutaneous bilirubinometry is a low-cost, noninvasive, and accurate screening test for hyperbilirubinemia with a high sensitivity specificity and positive predictive value (PPV).

Authors

Institutions

Publication Details

Journal
Annals of PIMS-Shaheed Zulfiqar Ali Bhutto Medical University
Published
2025-01-15
DOI
https://doi.org/10.48036/apims.v21i1.762
Citations
2
Primary Topic
Hematological disorders and diagnostics
Type
article
Field-Weighted Citation Impact
4.51
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Diagnostic Accuracy of Transcutaneous Bilirubinometry Versus Serum Bilirubin Level

Ayesha Waheed, Yasir Abbas, Safeer K. Mughal, Haider Sherazi et al.
2 citations
Annals of PIMS-Shaheed Zulfiqar Ali Bhutto Medical University
Hematological disorders and diagnostics
4.51
article

Diagnostic Accuracy of Transcutaneous Bilirubinometry Versus Serum Bilirubin Level

Ayesha Waheed, Yasir Abbas, Safeer K. Mughal, Haider Sherazi, Kinza Imran, Mustansir Ali
article en
2 citations

Abstract

Objective: To assess the diagnostic accuracy of transcutaneous bilirubinometry over serum bilirubin level in newborns with hyperbilirubinemia. Methodology: This cross-sectional study was conducted at Department of Neonatology, Children Hospital, PIMS Islamabad for six months from 01-09-20 to 28-02-21, included 150 jaundiced neonates of any gender over six months. Exclusion criteria included pre-treated hyperbilirubinemia, very premature or low birth weight infants, congenital abnormalities, or infants of jaundiced mothers. Post ethical clearance and parental consent, neonates were evaluated using the MBJ-20 transcutaneous bilirubinometer, with results compared to serum bilirubin levels. The study identified true and false diagnostic outcomes based on a ?12 mg/dL bilirubin threshold, signifying phototherapy initiation. Data analysis was performed with SPSS 26.0, calculating sensitivity, specificity, and accuracy, and determining diagnostic congruence through Pearson’s correlation and ROC AUC, with a significance threshold at p<0.05. Results: The mean serum bilirubin level and transcutaneous bilirubin levels were 14.28 ± 2.21 mg/dL and 14.52 ± 2.72 mg/dL respectively, with a mean difference of 0.24± 1.38 mg/dL (t(149) = 2.141, p = .034). This difference demonstrating a statistically significant overestimation compared to serum bilirubin levels. The correlation between transcutaneous and serum bilirubin levels was strong (r = 0.864, p < 0.0001). For the 12 mg/dL phototherapy threshold, the transcutaneous method showed high sensitivity (95.59%) and specificity (78.57%), with an overall test accuracy of 94%. The Area Under the Curve (AUC) for the receiver operating characteristic was 0.962, indicating high diagnostic precision. Conclusion: Transcutaneous bilirubinometry is a low-cost, noninvasive, and accurate screening test for hyperbilirubinemia with a high sensitivity specificity and positive predictive value (PPV).

Annals of PIMS-Shaheed Zulfiqar Ali Bhutto Medical UniversityVol. 21(1)
Riverbank Local Redevelopment Authority (US), Pakistan Institute of Medical Sciences (PK)
Openalex Percentile: Top 6%
Hematological disorders and diagnostics
4.51
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.