Performance Evaluation of Two Glucometers: RIGHTEST<sup>TM</sup> GM700S and RIGHTEST<sup>TM</sup> GM550 from Bionime

Glucometers are medical devices that provide glycemic monitoring. On-site verification of the performance of a glucometer is a requirement of ISO 22870. Thus, the objective of this study was to evaluate the performance of two glucometers (RIGHTESTTM GM700S and GM550). This is a prospective analytical study, carried out at the Biochemistry-Hematology Laboratory of the Fann University Hospital. The parameters studied were venous glycemia and capillary glycemia. The main evaluation parameters used were: repeatability, reproducibility, estimation of accuracy and comparison of methods. The performance criteria used were those of the French Society of Clinical Biology (SFBC) and the DIN EN ISO 15197: 2013 standard. The coefficients of variation of reproducibility (1.88%; 2.2%) and repeatability (1.93%; 1.8%) of the two glucometers complied with the SFBC criteria. The accuracy estimate showed a bias of −15% in accordance with the requirements of DIN EN ISO 15197:2013. The comparison of capillary glycemia values and those obtained with the reference automatons showed satisfactory linearity and good agreement. The glucometers showed acceptable performance compared to the accepted standards. They are suitable for self-monitoring of glycemia levels in diabetic patients.

Authors

Publication Details

Journal
Journal of Diabetes Mellitus
Published
2025-01-01
DOI
https://doi.org/10.4236/jdm.2025.153009
Primary Topic
Cancer, Hypoxia, and Metabolism
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance Evaluation of Two Glucometers: RIGHTESTTM GM700S and RIGHTESTTM GM550 from Bionime

Pape Matar Kandji, Nènè Oumou Kesso Barry, Papa Madièye Guéye, Najah Fatou Coly et al.
Journal of Diabetes Mellitus
Cancer, Hypoxia, and Metabolism
article

Performance Evaluation of Two Glucometers: RIGHTESTTM GM700S and RIGHTESTTM GM550 from Bionime

Pape Matar Kandji, Nènè Oumou Kesso Barry, Papa Madièye Guéye, Najah Fatou Coly, El Hadji Malick Ndour, Moustapha Djité, Ndèye Marème Thioune, S.M. Ndiaye, Fatou Gueye-Tall, Khoudia Diop
article en

Abstract

Glucometers are medical devices that provide glycemic monitoring. On-site verification of the performance of a glucometer is a requirement of ISO 22870. Thus, the objective of this study was to evaluate the performance of two glucometers (RIGHTESTTM GM700S and GM550). This is a prospective analytical study, carried out at the Biochemistry-Hematology Laboratory of the Fann University Hospital. The parameters studied were venous glycemia and capillary glycemia. The main evaluation parameters used were: repeatability, reproducibility, estimation of accuracy and comparison of methods. The performance criteria used were those of the French Society of Clinical Biology (SFBC) and the DIN EN ISO 15197: 2013 standard. The coefficients of variation of reproducibility (1.88%; 2.2%) and repeatability (1.93%; 1.8%) of the two glucometers complied with the SFBC criteria. The accuracy estimate showed a bias of −15% in accordance with the requirements of DIN EN ISO 15197:2013. The comparison of capillary glycemia values and those obtained with the reference automatons showed satisfactory linearity and good agreement. The glucometers showed acceptable performance compared to the accepted standards. They are suitable for self-monitoring of glycemia levels in diabetic patients.

Journal of Diabetes MellitusVol. 15(03)
Openalex Percentile: Top 75%
Cancer, Hypoxia, and Metabolism
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Performance Evaluation of Two Glucometers: RIGHTEST<sup>TM</sup> GM700S and RIGHTEST<sup>TM</sup> GM550 from Bionime — Pape Matar Kandji, Nènè Oumou Kesso Barry, et al. · Journal of Diabetes Mellitus (2025) | TGRS Research Map | TGRS