Advances in magnetic particle imaging: evaluating magnetic microspheres and optimized acquisition parameters for high sensitivity cell tracking

Introduction The sensitivity and resolution of magnetic particle imaging (MPI) depend on the choice of tracer and specific imaging parameters. For cell tracking applications with MPI, both the superparamagnetic iron oxide (SPIO) tracer and the cell labeling efficiency have a significant impact on MPI sensitivity and vary for different tracers. Methods This study compared three commercially available SPIO tracers (VivoTrax, Synomag-D and ProMag) and SPIO-labeled cells using magnetic particle relaxometry (MPR) and imaging. Further, the effect of imaging parameters (high and low gradient field strength and drive field amplitude) on MPI signal strength, resolution, and cell detection limits, was evaluated. Results The peak MPI signal measured by MPR was much higher for Synomag-D compared to VivoTrax and ProMag. However, the signal for intracellular Synomag-D was significantly reduced. In contrast, the signal for ProMag, a micron-sized iron oxide (MPIO) particle, was not significantly different for free and intracellular particles. The cellular iron loading was higher for ProMag compared to Synomag-D. The total MPI signal measured from images of free and intracellular SPIOs was highest for ProMag. Varying imaging parameters confirmed that a lower gradient field strength and higher drive field amplitude improved tracer and cellular sensitivity. Discussion These results, in addition to prior work from our lab, suggest that MPIOs are a good option for cell tracking with MPI. In conclusion, the evaluation of tracers by MPR is not sufficient to predict the performance of all SPIO tracers; in particular, not for larger, polymer-encapsulated iron particles such as ProMag, or for SPIO tracers internalized in cells. Improvements in MPI sensitivity through lower gradient field strength and higher drive field amplitudes are associated with a trade-off in image resolution.

Authors

Institutions

Publication Details

Journal
Frontiers in Imaging
Published
2025-07-24
DOI
https://doi.org/10.3389/fimag.2025.1610258
Primary Topic
Characterization and Applications of Magnetic Nanoparticles
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Advances in magnetic particle imaging: evaluating magnetic microspheres and optimized acquisition parameters for high sensitivity cell tracking

Paula J. Foster, Samantha Flood
Frontiers in Imaging
Characterization and Applications of Magnetic Nanoparticles
article

Advances in magnetic particle imaging: evaluating magnetic microspheres and optimized acquisition parameters for high sensitivity cell tracking

Paula J. Foster, Samantha Flood
article en

Abstract

Introduction The sensitivity and resolution of magnetic particle imaging (MPI) depend on the choice of tracer and specific imaging parameters. For cell tracking applications with MPI, both the superparamagnetic iron oxide (SPIO) tracer and the cell labeling efficiency have a significant impact on MPI sensitivity and vary for different tracers. Methods This study compared three commercially available SPIO tracers (VivoTrax, Synomag-D and ProMag) and SPIO-labeled cells using magnetic particle relaxometry (MPR) and imaging. Further, the effect of imaging parameters (high and low gradient field strength and drive field amplitude) on MPI signal strength, resolution, and cell detection limits, was evaluated. Results The peak MPI signal measured by MPR was much higher for Synomag-D compared to VivoTrax and ProMag. However, the signal for intracellular Synomag-D was significantly reduced. In contrast, the signal for ProMag, a micron-sized iron oxide (MPIO) particle, was not significantly different for free and intracellular particles. The cellular iron loading was higher for ProMag compared to Synomag-D. The total MPI signal measured from images of free and intracellular SPIOs was highest for ProMag. Varying imaging parameters confirmed that a lower gradient field strength and higher drive field amplitude improved tracer and cellular sensitivity. Discussion These results, in addition to prior work from our lab, suggest that MPIOs are a good option for cell tracking with MPI. In conclusion, the evaluation of tracers by MPR is not sufficient to predict the performance of all SPIO tracers; in particular, not for larger, polymer-encapsulated iron particles such as ProMag, or for SPIO tracers internalized in cells. Improvements in MPI sensitivity through lower gradient field strength and higher drive field amplitudes are associated with a trade-off in image resolution.

Frontiers in ImagingVol. 4
Western University (CA), Robarts Clinical Trials (CA)
Canadian Institutes of Health Research
Openalex Percentile: Top 83%
Characterization and Applications of Magnetic Nanoparticles
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.