Open Type Solid Photoacoustic Trace Gas Sensor with Multi-Pass Absorption Enhancement

A wavelength modulation solid photoacoustic spectroscopic (WM-SPAS) sensor enhanced with an open-type multi-pass cell (OMPC) is reported for highly sensitive detection of trace gases, especially suitable for highly corrosive and long optical-to-thermal (non-radiative) relaxation gaseous species. Such open configuration is quite different from traditional trace gas detection methods in that the separation design of the acoustic signal detector and gas absorption cavity avoids the adversely corrosive effect and reduces signal fluctuations caused by high flow rates. The modulated beam after the optical absorption by the target gas in the designed open-type multi-pass path is directed into a self-designed solid chamber, filled with carbon powder while the photoacoustic (PA) pressure signal is analyzed to yield the target gas concentration. By optimizing the incident beam angle, the OMPC achieves 96 reflections, yielding a 9.6 m optical path length enhancement. Using acetylene (C<sub>2</sub>H<sub>2</sub>) as a test sample and a DFB laser as the excitation source, this WM-SPAS sensor achieves sensitivity of 80 ppb and corresponding normalized noise equivalent absorption coefficient equal to 2.42 × 10<sup>-9</sup> cm<sup>-1</sup> W/Hz<sup>-1/2</sup> with 1 s time constant and modulation frequency as low as 39 Hz, which enables the sensor to detect gases with slow non-radiative relaxation. An Allan deviation analysis indicated the minimum detection limit could be further improved to 7 ppb at 100 s integration time. The response deviation of the PA signal under different flow rates was characterized by a coefficient of variation of 0.71‰. With its separate structure design, this newly developed PAS trace gas sensor offers unique advantages for open trace gas detection in high-flow and corrosive environments.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2025-07-24
DOI
https://doi.org/10.1021/acs.analchem.5c03412
Primary Topic
Spectroscopy and Laser Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Open Type Solid Photoacoustic Trace Gas Sensor with Multi-Pass Absorption Enhancement

Huiting Huan, Xuesen Xu, Lixian Liu, Huailiang Xu et al.
Analytical Chemistry
Spectroscopy and Laser Applications
article

Open Type Solid Photoacoustic Trace Gas Sensor with Multi-Pass Absorption Enhancement

Huiting Huan, Xuesen Xu, Lixian Liu, Huailiang Xu, Xueshi Zhang, J.Y Zhang, Baisong Chen, Yize Liang, Binxing Zhao, Jialiang Sun, Andreas Mandelis
article en

Abstract

A wavelength modulation solid photoacoustic spectroscopic (WM-SPAS) sensor enhanced with an open-type multi-pass cell (OMPC) is reported for highly sensitive detection of trace gases, especially suitable for highly corrosive and long optical-to-thermal (non-radiative) relaxation gaseous species. Such open configuration is quite different from traditional trace gas detection methods in that the separation design of the acoustic signal detector and gas absorption cavity avoids the adversely corrosive effect and reduces signal fluctuations caused by high flow rates. The modulated beam after the optical absorption by the target gas in the designed open-type multi-pass path is directed into a self-designed solid chamber, filled with carbon powder while the photoacoustic (PA) pressure signal is analyzed to yield the target gas concentration. By optimizing the incident beam angle, the OMPC achieves 96 reflections, yielding a 9.6 m optical path length enhancement. Using acetylene (C<sub>2</sub>H<sub>2</sub>) as a test sample and a DFB laser as the excitation source, this WM-SPAS sensor achieves sensitivity of 80 ppb and corresponding normalized noise equivalent absorption coefficient equal to 2.42 × 10<sup>-9</sup> cm<sup>-1</sup> W/Hz<sup>-1/2</sup> with 1 s time constant and modulation frequency as low as 39 Hz, which enables the sensor to detect gases with slow non-radiative relaxation. An Allan deviation analysis indicated the minimum detection limit could be further improved to 7 ppb at 100 s integration time. The response deviation of the PA signal under different flow rates was characterized by a coefficient of variation of 0.71‰. With its separate structure design, this newly developed PAS trace gas sensor offers unique advantages for open trace gas detection in high-flow and corrosive environments.

Analytical ChemistryVol. 97(30)
Xidian University (CN), University of Electronic Science and Technology of China (CN), China National Nuclear Corporation (CN), University of Toronto (CA), Integrated Optoelectronics (Norway) (NO), Institute for Advanced Study (DE), University of Chinese Academy of Sciences (CN)
National Natural Science Foundation of China, Natural Sciences and Engineering Research Council of Canada, Fundamental Research Funds for the Central Universities
Affordable and clean energy
Openalex Percentile: Top 84%
Spectroscopy and Laser Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.