Validation of Hemoglobin and Hematocrit Measurements from a Dialysis Machine Sensor Compared to Laboratory Analysis

<b>Background:</b> Continuous monitoring of hemoglobin (HB) and hematocrit (HCT) during hemodialysis could improve fluid management and patient safety. The Fresenius 5008 dialysis machine includes an ultrasound-based sensor that estimates HB and HCT values, though its accuracy compared to standard laboratory measurements remains unclear. <b>Methods:</b> This exploratory observational study assessed the agreement between sensor-derived and laboratory-derived HB and HCT values in 20 patients at the start of hemodiafiltration. A total of 177 paired blood samples were collected. <b>Results:</b> Sensor values significantly underestimated laboratory HB (9.61 vs. 11.31 g/dL) and HCT (27% vs. 34%) (<i>p</i> < 8 × 10<sup>-25</sup>). Correlations were strong for both parameters (HB: r = 0.788; HCT: r = 0.876). Regression analyses revealed consistent proportional bias. Applying a fixed correction of +1.69 g/dL for HB and +7.55% for HCT eliminated the statistical differences and reduced intercepts in regression models. Bland-Altman plots confirmed improved agreement post-correction. Albumin levels correlated modestly with error magnitude. <b>Conclusions:</b> HB and HCT values from the Fresenius 5008 sensor are strongly correlated with laboratory data but are systematically underestimated at treatment start, likely due to hemodilution. Applying fixed correction factors improves accuracy and supports the sensor's use for real-time monitoring.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2025-07-24
DOI
https://doi.org/10.3390/jcm14155242
Citations
1
Primary Topic
Central Venous Catheters and Hemodialysis
Type
article
Field-Weighted Citation Impact
2.52
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of Hemoglobin and Hematocrit Measurements from a Dialysis Machine Sensor Compared to Laboratory Analysis

Giulia Ligabue, Giulia Manzini, Licia Veronesi, Marco Ferrarini et al.
1 citations
Journal of Clinical Medicine
Central Venous Catheters and Hemodialysis
2.52
article

Validation of Hemoglobin and Hematocrit Measurements from a Dialysis Machine Sensor Compared to Laboratory Analysis

Giulia Ligabue, Giulia Manzini, Licia Veronesi, Marco Ferrarini, Grazia Maria Virzì, Gaetano Alfano, Valentina Di Pinto, Luigi Rovati, Silvia Giovanella, Niccolò Morisi, Gabriele Donati, Lucia Stipo, Fabio Olmeda
article en
1 citations

Abstract

<b>Background:</b> Continuous monitoring of hemoglobin (HB) and hematocrit (HCT) during hemodialysis could improve fluid management and patient safety. The Fresenius 5008 dialysis machine includes an ultrasound-based sensor that estimates HB and HCT values, though its accuracy compared to standard laboratory measurements remains unclear. <b>Methods:</b> This exploratory observational study assessed the agreement between sensor-derived and laboratory-derived HB and HCT values in 20 patients at the start of hemodiafiltration. A total of 177 paired blood samples were collected. <b>Results:</b> Sensor values significantly underestimated laboratory HB (9.61 vs. 11.31 g/dL) and HCT (27% vs. 34%) (<i>p</i> < 8 × 10<sup>-25</sup>). Correlations were strong for both parameters (HB: r = 0.788; HCT: r = 0.876). Regression analyses revealed consistent proportional bias. Applying a fixed correction of +1.69 g/dL for HB and +7.55% for HCT eliminated the statistical differences and reduced intercepts in regression models. Bland-Altman plots confirmed improved agreement post-correction. Albumin levels correlated modestly with error magnitude. <b>Conclusions:</b> HB and HCT values from the Fresenius 5008 sensor are strongly correlated with laboratory data but are systematically underestimated at treatment start, likely due to hemodilution. Applying fixed correction factors improves accuracy and supports the sensor's use for real-time monitoring.

Journal of Clinical MedicineVol. 14(15)
University of Modena and Reggio Emilia (IT), Azienda Ospedaliero-Universitaria di Modena (IT), Ospedale San Bortolo (IT), International Renal Research Institute of Vicenza (IT)
Openalex Percentile: Top 11%
Central Venous Catheters and Hemodialysis
2.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.