<i>Objective.</i>The CT dose index (CTDI) can be directly measured only for axial scan protocols; however, these days, helical protocols are applied in most cases, and sometimes axial protocols are not even available. The multiple scan average dose (MSAD) method can address this issue, but it is not widely adopted. This article aims to support the appropriateness of the MSAD method first by showing that it is measurable with the CTDI tools, and second by presenting a simple rule for scan pitch to avoid the main uncertainty of the measurement.<i>Approach.</i>Monte Carlo simulation was combined with its analytical postprocessing to accurately calculate scattering and speed up calculations to investigate several imaging and beam parameter combinations. The calculations were validated by measurements, and the model was used to examine the uncertainty of MSAD caused by an uncontrolled parameter.<i>Main results.</i>The widespread 150 mm long CTDI phantom and the 100 mm long ionization chamber are applicable for the MSAD method and reach80.8±0.2% efficiency in the estimation of CTDI<sub>vol</sub>, in contrast to78.0±0.3% efficiency of the CTDI<sub>100</sub>method. By utilizing the periodicity of the MSAD measurement, the uncertainty can be reduced to1.38±0.02% in contrast to26.85±0.16% in the worst case.<i>Significance.</i>The MSAD measurement with the CTDI setup is slightly more efficient than the CTDI<sub>100</sub>measurement protocol, so there is no need for a long phantom or farmer chamber to use MSAD in quality assurance (QA) measurements. The MSAD applies to both helical and axial scans; therefore, we recommend using direct MSAD measurement for QA in the clinical environment.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-07-24
DOI
https://doi.org/10.1088/1361-6560/adf40d
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Expansion of applicability of multiple scan average dose measurements for CT dosimetry

Dávid László Tárnoki, Csilla Pesznyák, Ákos Sudár, Ádám Domonkos Tárnoki et al.
Physics in Medicine and Biology
Radiation Dose and Imaging
article

Expansion of applicability of multiple scan average dose measurements for CT dosimetry

Dávid László Tárnoki, Csilla Pesznyák, Ákos Sudár, Ádám Domonkos Tárnoki, Tibor Major, Tamás Pócza, Richard Elek
article en

Abstract

<i>Objective.</i>The CT dose index (CTDI) can be directly measured only for axial scan protocols; however, these days, helical protocols are applied in most cases, and sometimes axial protocols are not even available. The multiple scan average dose (MSAD) method can address this issue, but it is not widely adopted. This article aims to support the appropriateness of the MSAD method first by showing that it is measurable with the CTDI tools, and second by presenting a simple rule for scan pitch to avoid the main uncertainty of the measurement.<i>Approach.</i>Monte Carlo simulation was combined with its analytical postprocessing to accurately calculate scattering and speed up calculations to investigate several imaging and beam parameter combinations. The calculations were validated by measurements, and the model was used to examine the uncertainty of MSAD caused by an uncontrolled parameter.<i>Main results.</i>The widespread 150 mm long CTDI phantom and the 100 mm long ionization chamber are applicable for the MSAD method and reach80.8±0.2% efficiency in the estimation of CTDI<sub>vol</sub>, in contrast to78.0±0.3% efficiency of the CTDI<sub>100</sub>method. By utilizing the periodicity of the MSAD measurement, the uncertainty can be reduced to1.38±0.02% in contrast to26.85±0.16% in the worst case.<i>Significance.</i>The MSAD measurement with the CTDI setup is slightly more efficient than the CTDI<sub>100</sub>measurement protocol, so there is no need for a long phantom or farmer chamber to use MSAD in quality assurance (QA) measurements. The MSAD applies to both helical and axial scans; therefore, we recommend using direct MSAD measurement for QA in the clinical environment.

Physics in Medicine and BiologyVol. 70(16)
Semmelweis University (HU), Budapest University of Technology and Economics (HU), National Institute of Oncology (HU)
Industry, innovation and infrastructure
Openalex Percentile: Top 73%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.