Utility and feasibility of intra-pocket mapping technique for optimal subcutaneous implantable cardioverter defibrillator implantation positioning for R-wave detection: a case series
These cases highlight the utility of intra-pocket mapping to optimize S-ICD positioning for R-wave detection in patients with challenging electrocardiogram characteristics. Posterior chest positioning may not be optimal for R-wave detection, particularly in patients with low R-wave and/or high T-wave amplitudes. Repositioning the device closer to the left ventricular apex improved sensing test results, supporting antero-inferior chest placement as a potential solution when posterior chest placement fails.
Authors
- Nobuhiko Ueda (ORCID: https://orcid.org/0000-0002-7862-3967)
- Yui Kitami
- Satoshi Oka (ORCID: https://orcid.org/0000-0001-6257-1421)
- Mitsuru Wada
- Taiju Watanabe
- KENGO FUKUSHIMA KUSANO (ORCID: https://orcid.org/0000-0002-5760-9285)
- Koji Ogawa (ORCID: https://orcid.org/0000-0002-9221-6815)
- Kohei Ishibashi (ORCID: https://orcid.org/0000-0001-7890-9554)
Institutions
- National Cerebral and Cardiovascular Center (JP)
Publication Details
- Journal
- European Heart Journal - Case Reports
- Published
- 2025-07-24
- DOI
- https://doi.org/10.1093/ehjcr/ytaf348
- Primary Topic
- Cardiac pacing and defibrillation studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00