Analysis of The Effect of Syringe Barrel Volume on Mechanical Failure During Injection
Background: Syringes are critical medical devices whose mechanical integrity is essential to ensure safety during injections. The cylinder volume can influence their mechanical performance. Objective: This study investigates how cylinder volume affects the mechanical integrity of syringes by analyzing von Mises stress, buckling risk, and overall safety during injection. Materials and Methods: Autodesk Inventor 2021 and Ansys 2019 R2 were used to analyze syringes with volumes of 1 ml, 3 ml, 5 ml, and 10 ml. Von Mises stress and buckling susceptibility were evaluated during simulated injections. Results: Smaller syringes showed higher stress levels: 14.030 MPa for the 1 ml syringe, 10.532 MPa for the 3 ml, and 7.0150 MPa for the 5 ml and 10 ml. The 5 ml syringe was the most susceptible to buckling, with a critical pressure of -0.1757 MPa. Despite this, all designs are safe under normal operating conditions. Conclusions: Cylinder volume affects mechanical performance, but with proper design, even smaller syringes can remain safe, enhancing the reliability of medical treatments.
Authors
- Krisdiyanto Acknowledgment
Institutions
- Muhammadiyah University of Yogyakarta (ID)
Publication Details
- Journal
- Gaceta Médica Boliviana
- Published
- 2025-06-30
- DOI
- https://doi.org/10.47993/gmb.v48i1.959
- Primary Topic
- Pharmacy and Medical Practices
- Type
- article
- Field-Weighted Citation Impact
- 0.00