The effect of inhalation delay on the aerodynamic particle size distribution of fluticasone propionate and ciclesonide using pMDI and valved holding chambers

Delayed inhalation when using pressurised metered-dose inhalers (pMDIs) for inhaled corticosteroids (ICS) is a common technique error. Valved holding chambers (VHCs) can mitigate its impact, but the effect of delays on drug delivery from suspension versus solution formulations remains poorly understood. We compared the doses of fluticasone propionate (FP) and ciclesonide (CIC) delivered to an anatomical adult throat model and a Next Generation Impactor as particles 1-5 μm and under 1 μm in diameter, using AeroChamber (AC), EasyChamber (EC), and OptiChamber Diamond (OD) VHCs with inhalation delays of 0, 1, 3, and 5 s. A breathing simulator was used to produce a single, adult-type inhalation. Throat deposition gradually decreased with longer inhalation for both FP and CIC, from an average of 3.3 %-2.2 % and 1.0 %-0.5 % of the label claim, respectively. Similarly, deposition of 1-5 μm particles declined from 30 % to 28 % for FP and from 33 % to 25 % for CIC. In contrast, deposition of particles smaller than 1 μm were relatively unaffected by inhalation delays. In conclusion, increasing the inhalation delay up to 5 s slightly reduced the respirable deposition of FP and CIC particles when using a VHC, but these reductions are unlikely to be clinically meaningful due to small differences in absolute dose. The smallest particles likely remained unaffected by the delay due to their low tendency to settle. Differences in performance between AC, EC, and OD during the delay were likely too minor to influence treatment outcomes.

Authors

Institutions

Publication Details

Journal
Respiratory Medicine
Published
2025-07-16
DOI
https://doi.org/10.1016/j.rmed.2025.108258
Primary Topic
Inhalation and Respiratory Drug Delivery
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The effect of inhalation delay on the aerodynamic particle size distribution of fluticasone propionate and ciclesonide using pMDI and valved holding chambers

Péter Csonka, Lauri Lehtimäki, Leon L. Csonka
Respiratory Medicine
Inhalation and Respiratory Drug Delivery
article

The effect of inhalation delay on the aerodynamic particle size distribution of fluticasone propionate and ciclesonide using pMDI and valved holding chambers

Péter Csonka, Lauri Lehtimäki, Leon L. Csonka
article en

Abstract

Delayed inhalation when using pressurised metered-dose inhalers (pMDIs) for inhaled corticosteroids (ICS) is a common technique error. Valved holding chambers (VHCs) can mitigate its impact, but the effect of delays on drug delivery from suspension versus solution formulations remains poorly understood. We compared the doses of fluticasone propionate (FP) and ciclesonide (CIC) delivered to an anatomical adult throat model and a Next Generation Impactor as particles 1-5 μm and under 1 μm in diameter, using AeroChamber (AC), EasyChamber (EC), and OptiChamber Diamond (OD) VHCs with inhalation delays of 0, 1, 3, and 5 s. A breathing simulator was used to produce a single, adult-type inhalation. Throat deposition gradually decreased with longer inhalation for both FP and CIC, from an average of 3.3 %-2.2 % and 1.0 %-0.5 % of the label claim, respectively. Similarly, deposition of 1-5 μm particles declined from 30 % to 28 % for FP and from 33 % to 25 % for CIC. In contrast, deposition of particles smaller than 1 μm were relatively unaffected by inhalation delays. In conclusion, increasing the inhalation delay up to 5 s slightly reduced the respirable deposition of FP and CIC particles when using a VHC, but these reductions are unlikely to be clinically meaningful due to small differences in absolute dose. The smallest particles likely remained unaffected by the delay due to their low tendency to settle. Differences in performance between AC, EC, and OD during the delay were likely too minor to influence treatment outcomes.

Respiratory MedicineVol. 247
Tampere University of Applied Sciences (FI), Tampere University Hospital (FI), Tampere University (FI)
Suomen Lääketieteen Säätiö, Jalmari ja Rauha Ahokkaan Säätiö, Tampereen Tuberkuloosisäätiö, Hengityssairauksien Tutkimussäätiö
Affordable and clean energy
Openalex Percentile: Top 72%
Inhalation and Respiratory Drug Delivery
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.